Vacuum pressure

The more sensitive the 

skin the less pressure 

should be used

Too much pressure is 

uncomfortable for the client 

and may lead to a greater 

peel than is needed.

The most sensitive skin is 

found around the eyes and 

on the neck below the chin.

We recommend a working 

pressure for sensitive skin 

of –2 PSI.

The rest of the face (and 

body) is less sensitive. 

Depending on the depth of 

peel and the condition of 

the skin a working pressure 

of –3 to –4 PSI is adequate

Beginning Treatment:

Once you have checked the following, the machine is ready to use:

 Tubing attached securely to Handpiece and to DermaVisage.

 Plastic treatment tip securely attached to Handpiece

 Fresh crystal container is filled to required level.

The first step is to set the working vacuum for the hand-piece.  

To do this, place your thumb over the hand-piece tip and turn on the 

machine. Keeping the hand-piece closed set the vacuum regulator to 

show –3 psi. (This is the standard setting for facial peeling) You are now 

ready to treat.

Treatment technique: 

Place a finger on the skin to one side of the area to be treated. This 

keeps the skin taut and hinders the skin from being pushed forward by 

the hand-piece. Keeping your finger placed here, apply the treatment tip 

gently to the skin and move it smoothly away from your finger.
It is better not to run the treatment-tip backwards and forwards over 

the same area of skin, but to start from the same position next to your 

finger and to move in one direction only. The treatment will otherwise be 

significantly more intensive.
The crystals can only flow when the system is completely closed, in 

other words when the treatment tip is in proper contact with the skin.
Always support the skin with one hand and 

do not exert any pressure 

on the skin with the hand-piece

. The vacuum itself ensures that the 

system is properly closed.
A hand-piece pushed too tightly against the skin will only cause irritation 

and reddening, with no positive effects on the intensity of the treatment.

General Treatment Guidelines

  Do not place the hand-piece onto the skin with too much pressure. 

This has no bearing on the peeling results. Too much physical 

pressure is uncomfortable for your client.

  Place a finger onto the skin and work away from this with the hand-

piece. This keeps the skin taught and hinders the skin form being 

pushed forward by the hand-piece.

  Avoid peeling in a circular motion. Peel in one direction only lifting the 

hand-piece from the skin at the end of each pass.

Содержание CC3040

Страница 1: ...DERMAVISAGE MAINTAINENCE AND TROUBLESHOOTING GUIDE INCORPORATING MODELS CC3040 CC3045 CC3050 Setting the standard in electrotherapy equipment...

Страница 2: ...Setting the standard in electrotherapy equipment Index Introduction page 3 Operating Instructions page 4 Looking After Your Machine page 8 Troubleshooting Guide page 10...

Страница 3: ...temperature range 0 to 50 C Permissible relative humidity range 20 to 75 Cleaning Before carrying out any cleaning procedure ensure that the mains lead is disconnected from the inlet at the rear of th...

Страница 4: ...t of the machine will illuminate red Press the Crystal Peel Selector Button and the button will change from red to green Diamond Peeling Selector Diamond Peeling Intensity Dial Diamond Peeling Directi...

Страница 5: ...to the appropriate outlets When these accessories are connected the corresponding Selector Buttons will illuminate red If the machine is left unattended with the Crystal Peel function switched on and...

Страница 6: ...being pushed forward by the hand piece Keeping your finger placed here apply the treatment tip gently to the skin and move it smoothly away from your finger It is better not to run the treatment tip...

Страница 7: ...vacuum pressure by three other means Running the treatment tip slowly over the skin The treatment is more intensive because the crystals work for a longer period on the same area of skin Running the t...

Страница 8: ...ccant Sachet The lid of the Fresh Crystal Container has a removable holder which contains a desiccant sachet The sachet contains orange beads of desiccant which keep the air and therefore the crystals...

Страница 9: ...the Used Crystal Container When you are satisfied the tubes are clear release your fingers from the remaining outlets and turn the Vacuum Intensity back down Replace the container back into the base E...

Страница 10: ...the longer metal body should be connected via the shorter length of tubing to the connector in the centre of the machine Important Please note the following procedure If the lid has been fitted in rev...

Страница 11: ...K Unless otherwise stated please ensure the hand piece is connected properly the vacuum control on the front of the panel is turned fully clockwise to maximum and both containers are empty Did The Mac...

Страница 12: ...he black hand piece is pushed fully into outlet 2 Check to see if the tubing that connects the hand piece to the machine is damaged or split and that both tubes are pushed fully into the hand piece Ch...

Страница 13: ...psi and should drop to 0 5 or no display when you remove your finger from the Treatment Tip or the end of the tubing Did The Machine Indicate Full Vacuum YES Switch off and refit the fresh crystal con...

Страница 14: ...14...

Отзывы: