Dermavisage 

Operating Instructions

Installation And Preparation For Use

Unpack the unit before first use, and check you have everything listed on 

the Packing List.  

The unit is shipped to you ready for use. 

The Handpiece:  

This consists of the following components:  - Plastic Treatment Tip - Crystal 

Nozzle – Handpiece Body – Tubing (Clear & Black)
Before use, the tubing will need to be attached to the front of the unit.   
When fixing the Handpiece and tubing to the unit, ensure Tube 1 and 

Tube 2 are lined up with the corresponding outlet on the front of the 

DermaVisage, and push 

firmly

 into place using a slight twisting action. 

Turning the DermaVisage on: 

Turn the DermaVisage on via the Green On/Off switch at the back of the 

machine.  The Crystal Peel Selector Button on the front of the machine will 

illuminate red.  Press the Crystal Peel Selector Button, and the button will 

change from red to green.  

Diamond Peeling Selector 

Diamond Peeling  

Intensity Dial

Diamond Peeling 

Directional Rotation 

Switch

Outlet for Diamond Peeling 

Handpiece

Colour Therapy Selector Button

Outlet for Colour  

Therapy Applicator

Vacuum Suction 

Intensity Level

Crystal Peeling 

Selector

Crystal Peeling Vacuum 

Suction Intensity Dial

Outlet for  

Tube 1

Outlet for  

Tube 2

Crystal Dispensing Canisters



Treatment Tips: 

To operate, a plastic 

treatment tip must be 

attached to the working 

end of the Handpiece. To 

attach, simply push the tip 

firmly into place using a 

slight twisting action. These 

should be disposed of after 

each treatment. 

The Crystal Peeling (8), 

Diamond Peeling (1) and 

Colour Therapy (5) Selector 

Buttons all incorporate a 

“Switching Delay”

.  When 

pressed the buttons will 

only change from On to Off 

(and vice versa) if 3 seconds 

elapse between each press 

of the button.

Содержание CC3040

Страница 1: ...DERMAVISAGE MAINTAINENCE AND TROUBLESHOOTING GUIDE INCORPORATING MODELS CC3040 CC3045 CC3050 Setting the standard in electrotherapy equipment...

Страница 2: ...Setting the standard in electrotherapy equipment Index Introduction page 3 Operating Instructions page 4 Looking After Your Machine page 8 Troubleshooting Guide page 10...

Страница 3: ...temperature range 0 to 50 C Permissible relative humidity range 20 to 75 Cleaning Before carrying out any cleaning procedure ensure that the mains lead is disconnected from the inlet at the rear of th...

Страница 4: ...t of the machine will illuminate red Press the Crystal Peel Selector Button and the button will change from red to green Diamond Peeling Selector Diamond Peeling Intensity Dial Diamond Peeling Directi...

Страница 5: ...to the appropriate outlets When these accessories are connected the corresponding Selector Buttons will illuminate red If the machine is left unattended with the Crystal Peel function switched on and...

Страница 6: ...being pushed forward by the hand piece Keeping your finger placed here apply the treatment tip gently to the skin and move it smoothly away from your finger It is better not to run the treatment tip...

Страница 7: ...vacuum pressure by three other means Running the treatment tip slowly over the skin The treatment is more intensive because the crystals work for a longer period on the same area of skin Running the t...

Страница 8: ...ccant Sachet The lid of the Fresh Crystal Container has a removable holder which contains a desiccant sachet The sachet contains orange beads of desiccant which keep the air and therefore the crystals...

Страница 9: ...the Used Crystal Container When you are satisfied the tubes are clear release your fingers from the remaining outlets and turn the Vacuum Intensity back down Replace the container back into the base E...

Страница 10: ...the longer metal body should be connected via the shorter length of tubing to the connector in the centre of the machine Important Please note the following procedure If the lid has been fitted in rev...

Страница 11: ...K Unless otherwise stated please ensure the hand piece is connected properly the vacuum control on the front of the panel is turned fully clockwise to maximum and both containers are empty Did The Mac...

Страница 12: ...he black hand piece is pushed fully into outlet 2 Check to see if the tubing that connects the hand piece to the machine is damaged or split and that both tubes are pushed fully into the hand piece Ch...

Страница 13: ...psi and should drop to 0 5 or no display when you remove your finger from the Treatment Tip or the end of the tubing Did The Machine Indicate Full Vacuum YES Switch off and refit the fresh crystal con...

Страница 14: ...14...

Отзывы: