   U-2S2303-30121-G V1.0   Cari Chair

OPERATION MANUAL

Resistance Test

Test Level as per

IEC 60601

Level of Compliance

Electromagnetic 

Environment

Power frequency(50, 60 
Hz) magnetic field IEC 
61000-4-8

3 A/m

3 A/m

SR321 power frequency 
m a g n e t i c   f i e l d s 
should be at levels 
c h a r a c t e r i s t i c   o f   a 
typical location in a 
typical commercial or 
hospital environment.

NOTE:

 UT is the a.c. mains voltage prior to application of the test level.

 

Resistance Test

Test Level as per

IEC 60601

Level of 

Compliance

Electromagnetic Environment

Conducted RF
IEC 61000-4-6

3 Vrms
150 kHz to 80 
MHz

3 Vrms

Portable and mobile RF communications 
equipment should be used no closer to any 
part of SR321 including cables, than the 
recommended separation distance calculated 
from the equation applicable to the frequency 
of the transmitter.
Recommended distances:

d = 1.2√P
d = 1.2√P  80 MHz to 800 MHz
d = 2.3 √P  800 MHz to 2.5 GHz

Where P is the maximum output power rating 
of the transmitter in watts (W) according 
to the transmitter manufacturer and d is the 
recommended separation distance in mattress 
(m). 

Field strengths from fixed RF transmitters, as 

determined by an electromagnetic site survey 
a, should be less than the compliance level in 
each frequency range b. 
Interference may occur in the vicinity of 
equipment marked with the following 
symbol:

Radiated high-
frequency
phenomena
IEC 61000-4-3

3 V/m
80 MHz to 2.5 
GHz

3 V/m

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land 

mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with 

accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey 
should be considered. If the measured field strength in the location in which SR321 is used exceeds the applicable 

RF compliance level above, SR321 should be observed to verify normal operation. If abnormal performance is 
observed, additional measures may be necessary, such as re-orienting or relocating SR321

b

 

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

NOTE  

At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE  

These guidelines may not apply in all situations. Electromagnetic propagation is 

affected by absorption and reflection from structures, objects and people.

Содержание Cari Chair

Страница 1: ...Cari Chair Operation Manual Anti Decubitus Air Alternating Pressure Cushion Replacement System...

Страница 2: ...ESE INSTRUCTIONS CAREFULLY BEFORE USE AND OBSERVE THE SAFETY INSTRUCTIONS AND THE REQUIREMENTS FOR THE OPERATION AND MAINTENANCE OF THE DEVICE Use genuine Carilex components are essential for optimal...

Страница 3: ...stem Package 5 5 0 Features 6 6 0 Mattress and Power Unit Installations 8 7 0 Program Settings 8 8 0 Cleaning Instruction 9 Section Description Page 9 0 Routine Maintenance 9 10 0 Troubleshooting 10 1...

Страница 4: ...of pressure ulcer This device is intended to prevent pressure ulcers by facilitating blood circulation and decreasing pressure of each tissue s contact area If there s any question please always consu...

Страница 5: ...ients to avoid of cross contamination 8 Be sure to verify the patient weight does not exceed system weight capacity The maximum weight capacity of this device is 113 kg 250 lb 9 Charging the power uni...

Страница 6: ...to ensure maximum pressure relief The yellow LED indicates the device is monitoring the cushion s internal pressure while the compressor is inflating Battery Indicator The green LED bar indicates the...

Страница 7: ...will be triggered to notify medical staff when the cushion has insufficient internal pressure The power unit will automatically turn off if the low pressure indicator stays for 15 minutes Panel Lock...

Страница 8: ...ng to the next step 3 Secure the power unit to the rear of the chair by using dedicated carry bag 4 Firmly connect the air hose couplings to the quick release couplers on the power unit s air outlet 5...

Страница 9: ...lor might run insignificantly 4 DO NOT use electric or tumble dryers 5 DO NOT iron WARNING Always unplug the power unit before cleaning Routine cleaning of power unit can be done by wiping down with d...

Страница 10: ...rmly into place Verify if patient weight setting is correct Increase or decrease weight setting until appropriate pressure is reached Inspect air filter for dust Clean or replace air filter Lift cushi...

Страница 11: ...roduct on which the production lot number has been removed or defaced on which repair has been attempted by any person or agency not authorized by Carilex or if in the sole opinion of Carilex that the...

Страница 12: ...n 180 x 50 x 14 cm 26 air cells 70D nylon with TPU lamination Coverlet Two way stretch 40 polyurethane 60 polyester Wheel Chair Cushion Set Cushion dimension 17 7 x 17 7 x 3 5 in 45 x 45 x 9 cm 10 air...

Страница 13: ...ir Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast transient burst IEC 61000 4 4 2kV for power...

Страница 14: ...ress m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vic...

Отзывы: