ANTI-DECUBITUS AIR ALTERNATING PRESSURE CUSHION REPLACEMENT SYSTEM
9
English
10.0 Troubleshooting
Problem
Inspection Procedure
Possible Solutions
1. Power unit
d o e s n o t
function.
Check if it is “Low Battery”
Connect the adapter to the power outlet
to charge the power unit.
Check if the rocker switch is in ON
position.
Turn rocker switch to ON position.
Check if both of orange LED is
blinking and audible indicator sounds
simultaneously.
Please contact your authorized local
distributor for assistance.
2. Low pressure
LED is flashing
during operation.
Check if there is leakage in air tubes or
air cells.
Please contact your authorized local
distributor for assistance.
3 . P o w e r U n i t
is working, but
cushion is not
inflating.
Check if cushion’s air hose couplings
are properly connected to power unit’s
quick release couplers.
Secure air hose couplings firmly into
place.
Verify if patient weight setting is
correct.
Increase or decrease weight setting
until appropriate pressure is reached.
Inspect air filter for dust.
Clean or replace air filter.
Lift cushion coverlet up to check if air
cells are connected correctly.
Make sure all air cells are properly
linked to air supply
Lift cushion coverlet up to check if air
tubes are kinked or obstructed.
Check and adjust air tubes positions.
Check if air cells are cut or cracked.
Replace with genuine spare parts or
contact your local distributor.
4. Alternating and
static setting is
not available.
Check if the panel lock
Simply press this Multi-Function
Button for 3 seconds to unlock the
panel.
Possible control failure
Please contact your authorized local
distributor for assistance.
5. Patient is not
getting pressure
r e l i e f d u e t o
system failure
( r e d d e n i n g o f
skin)
Contact your physician and/ or nursing
service immediately.
If power unit does not respond to the possible solution, please contact your authorized local
distributor for assistance.