RVG 142, RVG 5200 and RVG 6200 Safety, Regulatory User Guide (SM847) Ed06 5
Warning and Safety Instructions
WARNINGS:
Device
•
Read and understand this safety information before using RVG 142, RVG 5200
and RVG 6200.
•
You are responsible for the operation and maintenance of this device. Only
legally qualified persons can operate this device. They MUST have training to
use the device. When necessary, have a trained authorized service technician
carry out inspection and maintenance operations.
•
Install this device in an X-ray room that complies with current installation
standards. From this location, you must be able to maintain visual or audio
communication with the patient and be able to access the Acquisition Interface
module during exposure.
•
Do NOT operate the device if there is the threat of an earthquake. Following an
earthquake, ensure that the device is operating satisfactorily before using it
again. Failure to observe this precaution may expose patients to hazards.
•
X-ray equipment is hazardous to patients and the operator if you do not observe
the exposure safety factors and operating instructions.
•
To dispose of the device or its components, contact a service technician.
•
No modification of this device is allowed.
•
Do NOT use this device in conjunction with oxygen-rich environments. This
device is not intended for use with flammable anesthetics or flammable agents.
•
It is not recommended to use accessories other than those specified in this
document and sold by Carestream Dental.
•
Single-use disposable protective hygienic sleeves must cover the RVG sensor
before placing it in the patient’s mouth.
•
To prevent cross-contamination, use a new hygienic barrier for each new
patient and disinfect the RVG sensor.
•
To ensure the best quality images, you must connect the RVG sensor to a USB
port on the BACK of the workstation.
•
RVG Connect: the management and the allocation of the static IP address
must be done by a network administrator.
•
RVG Connect: For the smooth functioning of the image recovery mechanism,
do NOT switch off the RVG Connect unit and do NOT disconnect the ethernet
cable from the RVG Connect unit.
•
Place the sensor in the patient mouth ONLY if the RVG icon is green on the
imaging software.
•
If you find that the RVG sensor is too hot, disconnect the RVG sensor and
contact an authorized service technician.
Computer
•
RVG 5200 or RVG 6200 or RVG 142 shall be supplied by Low Voltage-Low Power
(<15W) source separated from the mains by DI/RI Isolation (Double isolation /
Reinforced isolation). The USB port of a computer or the hub is separated from
the mains by DI/RI Isolation.
•
See your computer installation guide for details of the data processing system
and screen. Leave a sufficient amount of clear space around the CPU to ensure
that it is properly ventilated.
•
Always use Microsoft Windows Update to make sure that the latest security
patches are correctly installed.