CS 7600_Safety, Regulatory & Technical Specifications (8J4070)_Ed02
3
•
The appliance coupler/inlet of the flexible power cord is used as a
disconnecting device from the mains.
•
To dispose of the device or its components, contact a service
technician.
•
No modification of this equipment is allowed.
•
DO NOT block air circulation around the unit. Always maintain at
least 15 cm clearance around the unit to prevent overheating and
damage to the system.
•
The scanner should be positioned so that there is always easy access
to the mains power supply socket.
Imaging Plates:
•
To prevent damage to the imaging plates and the possibility of image
artifacts, avoid contact between the imaging plates and the following
materials/solutions/solvents: Isopropyl alcohol, hydrogen peroxide
and other peroxides, citrus-based cleaners, hand lotions and
water-less hand sanitizers, as well as surfactants and lubricants.
•
The imaging plate contains Barium and should be considered
hazardous or special waste in specific conditions at the end of its
useful service life. For disposal or recycling information, contact
your local authorities.
•
Do not soak the imaging plate in any cleaning or disinfecting
solutions. Do not autoclave; autoclaved imaging plates must be
discarded.
•
Imaging plates should be stored in their original packing or storage
box when they are not in use. Always store imaging plates in a dark
and dry place.
•
Do not expose the imaging plates to light for long periods as this can
have a degrading effect.
•
Do not store imaging plates in hot or moist conditions.
•
Do not fold, crease, or bend the imaging plates.
•
Avoid touching the imaging side of the imaging plates and be careful
not to drag the imaging side of the imaging plate across any surface
as this will damage the imaging plate.
•
Do not leave imaging plates where they can become damaged by
liquid or chemical spills.
•
Read and follow the instructions in Material Safety Data Sheets
(MSDS) for the CS Screen Cleaner.