CS 3800 Family User and Installation Guide (TA3975)_Ed02
17
CS 3800 Family Shade Calibration Unit Overview
After 50 hours of scanning (approximately 13 days), use the shade calibration unit to recalibrate the
scanner.
Figure 18
Shade Calibration Unit
Follow these recommendations:
Keep the cap on the calibration unit until you are ready to use it.
Always install a cleaned and sterilized tip on the scanner before attaching the shade
calibration unit.
Attach the shade calibration unit to the normal tip using the Normal opening.
Attach the shade calibration unit to the side tip using the Side opening.
Push the shade calibration unit to the end when attaching to the tip.
Note:
The posterior tip
cannot
be used for Shade Matching.
Do not touch the gray card in the shade calibration unit or expose it to liquids.
Store the calibration unit away from light, heat, and moisture.
Order a new shade calibration unit if the gray card quality check fails when performing a
calibration or if the expiration date on the shade calibration unit is approaching.
See the
CS ScanFlow User Guide
for more information.