![CareFusion Alaris TIVA Скачать руководство пользователя страница 20](http://html.mh-extra.com/html/carefusion/alaris-tiva/alaris-tiva_user-manual_2557176020.webp)
1000DF00330 Issue 4
19/34
Configured Options
This menu comprises a list of options which are configurable by the user.
1. Turn the pump
OFF
.
2. Whilst holding down the
b
button turn the pump
ON
.
3. The main display will show
000
. Enter the access code for Configured Options using the
f
keys, pressing
NEXT
to move through
the digits. A full list of access codes can be found in the Technical Service Manual.
4. When the complete code shows on screen, press
OK
to enter. The Configured Options menu will be displayed.
1. Select
GENERAL OPTIONS
from the menu using the
f
keys and press the
OK
softkey.
2. Select the option you wish to enable/disable or adjust and press the
MODIFY
softkey.
3. When all the desired modifications have been carried out press the
QUIT
softkey.
4. Either select the next configuration option from the menu or turn the pump
OFF
, returning it to operation as required.
NURSE CALL FITTED
Enables Nurse Call (hardware option).
NURSE CALL INVERT
When enabled, the nurse call output is inverted.
RS232 SELECTED
Sets the pump's communications to use RS232 (hardware option).
NEOI WARNING
Sets the Near End Of Infusion warning time, as time left to End Of Infusion.
EOI POINT
Sets the End Of Infusion point.
KVO AT EOI
When enabled the pump will switch to running at the KVO rate when EOI is reached.
KVO RATE
Sets the Keep Vein Open (KVO) rate at which the pump will operate if KVO at EOI is enabled.
BACK OFF
When enabled the motor will reverse to relieve line pressure when an occlusion occurs.
AC FAIL
When enabled the AC Power Failure Alarm will sound if the AC power is disconnected.
PRESSURE DISPLAY
Enables / disables the Pressure Icon on the main display.
PRESSURE DEFAULT
Sets the default occlusion alarm level.
WEIGHT
Sets up default patient weight, for TIVA mode only.
PURGE RATE
Sets the purge rate.
PURGE VOLUME LIMIT
Sets the maximum permissible purge volume.
PURGE SYRINGE
Prompt to purge syringe after confirmation.
HANDS FREE BOLUS
Enables / disables the hands free bolus feature.
DEFAULT BOLUS VOL
Sets the default hands free bolus volume, for no drug mode only.
DEFAULT BOLUS RATE
Sets the default bolus rate.
MANUAL BOLUS
Volume infused will be increased if plunger is manually moved in and syringe remains confirmed.
CALL BACK TIME
Adjusts the time for the pump to sound the call back alarm.
EVENT LOG DISPLAY
Enables / disables the event log.
BATTERY ICON
Enables / disables the Battery Icon on the main display.
AUDIO VOLUME
Sets the alarm volume of the pump at high, medium or low.
AUTO NIGHT MODE
Backlight dims between hours 21:00 and 06:00.
General Options