Basic Features (Continued)
Pressure Level
1. To check and adjust the pressure level press the
e
button. A bar graph will be displayed showing the pressure alarm level and the
current pressure level.
2. Press the
f
keys to increase or decrease the alarm level. The new level will be indicated on the display.
3. Press
OK
to exit the screen.
A
The interpretation of pressure readings and occlusion alarms are the responsibility of the clinician depending on
the specific application.
During a bolus or induction phase the pressure level is set to maximum. The level remains at maximum for 10
seconds after the end of the phase.
If Rate Titration is
enabled
the rate can be adjusted
while infusing
:
1. Select the new rate using the
f
keys.
The message < START TO CONFIRM > will flash on screen and pump continues to infuse at the original rate.
2. Press the
b
button
to confirm the new infusion rate and start infusing at the new rate.
If Rate Titration is
disabled
the rate can only be adjusted
whilst on hold:
1. Press the
h
button to put the pump on hold.
2. Select the new rate using the
f
keys.
3. Press the
b
button
to start infusing at the new rate.
Rate Titration
1. To clear the induction, press the
h
button.
2. The
CLEAR INDUCTION
prompt is shown.
l
To clear the induction press the
YES
softkey.
l
Answering
NO
to the
CLEAR
INDUCTION
prompt will leave the pump in the
ON HOLD
state; the induction can continue by
pressing the
b
button.
Clear Induction (TIVA Mode only)
Clear Volume
This option enables the volume infused to be cleared.
1. Press the
VOLUME
softkey to display the
CLEAR VOLUME
option.
2. Press the
YES
softkey to clear the volume. Press the
NO
softkey to retain the volume.
Selecting YES resets the volume infused in the 24H LOG option.
To set doserate or flowrate in precise increments it may be necessary to switch between the rate adjust options
SET BY DOSERATE
and
SET BY ml/h
. An arrow to the left of the rate display shows the rate changed when the
f
keys are used to increase/decrease the
infusion rate.
To set a doserate precisely the arrow must be pointing to the doserate (mg/kg/h); the flowrate will be calculated from the doserate.
To precisely set a flowrate the arrow must be pointing to flowrate (ml/h); the doserate will be calculated from the flowrate.
Selecting the Set By ml/h Option
1. Whilst the pump is infusing, press the
d
button to access the options menu.
2. Select the
SET BY ml/h
option using the
f
keys and press the
OK
softkey indicated on the screen. This will select the set by
flowrate option, the arrow on the display will automatically select the flowrate, the flowrate can be adjusted if necessary.
Selecting the Set By Doserate Option
1. Whilst the pump is infusing, press the
d
button to access the options menu.
2. Select the
SET BY DOSERATE
option using the
f
keys and press the
OK
softkey indicated on the screen. This will select the set
by doserate option, the arrow on the display will automatically select the doserate, the doserate can be adjusted if necessary.
? Set by Doserate / Set by ml/h (TIVA Mode only)