9
Digi Blood Pressure Monitor -
DL08146
Measuring procedure:
After the cuff has been appropriately positioned, the measurement can begin:
1. Press the START/STOP button, all symbols appear on the display, you can hear 2 short
beeps after 0 flashes for 2 seconds, then the pump begins to inflate the cuff, the rising
pressure in the cuff is shown on the display.
2. After the suitable pressure has been reached, the pump stops and the pressure
gradually falls. The cuff pressure is displayed. In case that the inflation is not sufficient,
the device automatically re-inflates to a higher pressure.
3. When the device detects the signal, the heart symbol on the display starts to flash,
you can hear the beep for every heartbeat once the heartbeat signal is detected.
4. When the measurement has been completed, you can hear a long beep, in the
meantime, the systolic, diastolic and pulse rate will appear on the display.
5. The measurement readings remain on the display until you switch off the device. If no
button is pressed for a period of 3 minutes, the device switches itself off order to save
power.
Discontinuing a measurement
Memory-recall of measurements
Memory-recall of measurements
If it is necessary to interrupt a blood pressure measurement for any reason (eg. the patient feels
unwell) the START/STOP button can be pressed at any time. The device immediately decrease the cuff
pressure automatically.
This blood pressure monitor automatically stores 2*120 sets measurements value, the oldest record
will be replaced by the latest measurement value when more than 120 sets each user.
Press the button when power off, the last measurement value will be shown, and subsequent
measurements can be displayed one after the other by pressing the button each time.
Note: It will read in kPa as unit when you choose kPa, the symbol will be displayed along with the
reading if irregular heartbeat is detected during the measurement.
M