4.13. 

Create a new program

 

Compact4 device can create and save 50 new user pro-
grams. To create a new program, select “Programs” from 
the main menu and then the option “create a program”.

4.13.1. Name the new program

The first step after clicking “create a program” is to name the 
new program. You can enter it using the keyboard displayed 
on the screen. Confirm the name by pressing “OK” button. 

4.13.2. Setting parameters: time, speed, interval, hold

In the next step, set the parameters for the new program, 
such as time, speed, hold, interval.

Enter created program name

q

w

e

r

t

y

u

i

o

p

a

s

d

f

g

h

j

k

l

123

z

x

c

v

b

n

m

del

OK

space

1

Lymphoedema I

mmHg

min

level

220

50

5

3

3

Time

Pressure

Speed

Interval Hold

Gradient

1

2

3

4

2

2

Interval Hold

Gradient

1

2

3

4

PU1. THE NEW PROGRAM

1

2

SAVE

min

level

5

1

Speed

Time

2

2

Interval Hold

Gradient

1

2

3

4

PU1. THE NEW PROGRAM

1

SAVE

min

level

5

1

Speed

Time

1

2

1

2

Содержание COMPACT 4

Страница 1: ...Lymph Drainage Therapy Device Pressotherapy Device User Manual UM_EN_CPCOMPACT4_01...

Страница 2: ...figuration 13 4 1 Main screen 13 4 2 Work mode setting with provided programs 13 4 2 2 Provided programs 13 4 2 3 User programs 14 4 2 4 Favorite programs 14 4 2 5 Create a program 14 4 3 Pressure lev...

Страница 3: ...1 Energy security 31 5 2 Possibility of injuries 32 5 3 Safe use and maintenance 33 5 4 Correct use of the cuffs 33 6 Technical parameters 34 7 Contents of the package 35 8 Additional accessories 36 9...

Страница 4: ...tients athletes and all other users CarePump devices are used for regeneration relaxation and improving your well being The device works very similar to manual massage which can be performed on your o...

Страница 5: ...with the cuffs Cuffs used on individual parts of the body arms legs etc By sequential inflation of the cuff chambers a part of the body where the cuff is placed is being compressed to allow achieving...

Страница 6: ...uses a sequential pressure effect which turns into a rhythmic wave Lymphatic drainage massage is pos itively perceived as a very pleasant and effective way to re generate and relax The patient feels r...

Страница 7: ...ebitis skin diseases moles lymphangitis dermatitis wounds arthritis high body temperature thick external visible varicose veins cardiac rhythm disturbances arrhythmia blood pressure disorders myocardi...

Страница 8: ...be fore starting work with the device 2 1 CarePump Compact4 device front view main power switch input to connect the air ducts control keyboard touch screen display 2 2 CarePump Compact4 device back...

Страница 9: ...dequate for its weight Remember not to place the device too close to the wall because the holes of cooling fans are located on the back of the case Do not place on or near the device other electrical...

Страница 10: ...er end of the cord into a power socket Attention Use the supplied power cord or one that complies with electrical standards Make sure that the voltage in the installation is 100 240V 50 60Hz If the ca...

Страница 11: ...ing the air duct or the splitter pay attention to the notch which indicates proper connection of them To connect the air duct push it into the correct input until you hear a click of both side button...

Страница 12: ...rs of the cuff with correct order The proper con nection is indicated by a dark gray plug The separate ends should be connected in following order from the longest to the shortest one Correctly plugge...

Страница 13: ...hen the main screen view will appear where you can set basic treatment options 4 2 2 Provided programs 4 2 2 1 Setting using the keyboard Press any key on the keyboard to activate the icon named Progr...

Страница 14: ...s Create new program 4 2 2 2 Setting using the touch screen Press the icon Programs in the upper left corner of the display Then following screen will appear where you can select provided programs use...

Страница 15: ...it by pressing OK button Attention Pressure as a parameter is not defined for provided pro grams It should be adapted to the needs of each patient based on the patient s blood pressure and his feeling...

Страница 16: ...lling list with values 5 105 minutes will be displayed on the screen Values on the list can be displayed with an ac curacy of 1 5 and 10 You can decide on the accuracy of the displayed values by click...

Страница 17: ...A in the upper right corner of the display and then select one of the body areas highlighted in white B When you press arm leg hip and abdomen area or hip and buttock area the selected body part will...

Страница 18: ...amber press the corresponding area on the graphic image showing the cuff A scrolling list with pressure values in mmHg will be displayed on the screen Once the required value of the pressure is highli...

Страница 19: ...ers will drop gradually starting from the value set in starting pressure up to the value specified in fi nal pressure The pressure drop can be set in the pressure unit mmHg or as a percentage value It...

Страница 20: ...Hold function The parameter Interval A allows you to precisely de fine the time between subsequent cycles of filling the chambers The parameter Hold B allows you to pre cisely set the time of keeping...

Страница 21: ...t the appropriate language press lan guage and select the desired option from the list The ac tive language will be marked with symbol SETTINGS v2 3 0 Language ENG Sounds Password management Other CHA...

Страница 22: ...ed confirm it by pressing OK button Device ready turning on off the device standby sound after start up and selecting the volume of it End of therapy turning on off the acoustic sig nals after the end...

Страница 23: ...click on the icon with crossed eye next to the asterisks symbolizing the password The manufacturer recommends saving the password in a safe place inaccessible for unauthorized persons If you set a pa...

Страница 24: ...ction that allows you to select whether after starting the device the home screen will be the main screen or a list of available programs Button brightness function that allows you to set the keyboard...

Страница 25: ...shown by the blinking chamber on the graphic image of the cuff and the elapsing time To temporarily stop the treatment press pause icon B The treatment will stop and the treatment time will be paused...

Страница 26: ...HOEDEMA I The course of the treatment Single filling of consecutive chambers while keeping the pressure in previously inflated chambers Time 50 minutes Speed 5 Interval 3 sec Hold 3 sec Gradient 1 mmH...

Страница 27: ...tes during pumping of the third one Time 60 minutes Speed 5 Interval 3 sec Hold 3 sec Gradient 2 mmHg 4 SPORT POST WORKOUT RECOVERY The course of the treatment Double ascending wave Simultaneous infla...

Страница 28: ...dle one keeps the pres sure The sequence is repeated for the consecutive pumped chambers Always one chamber keeps the pressure Time 60 minutes Speed 5 Interval 2 sec Hold 2 sec Gradient 0 mmHg 6 ALL C...

Страница 29: ...yboard displayed on the screen Confirm the name by pressing OK button 4 13 2 Setting parameters time speed interval hold In the next step set the parameters for the new program such as time speed hold...

Страница 30: ...marked with a white dot A To return to the view of the previous sequence use button 20 sequences can be set for each new program The program can be added to the list of favorite programs during creati...

Страница 31: ...network are consistent with the data provided on the rating plate at the bottom of the case The manufacturer recommends connecting the de vices to an installation equipped with a residual current dev...

Страница 32: ...e adapt ed to adults Using them on younger patients may lead to permanent injuries with an implanted artificial heart and other heart prostheses with metal or joint implants as the procedure may cause...

Страница 33: ...5 m the floor should be made of non conductive materials the ground on which the device is located should be stable made of non conductive and non slip materials the device should not be exposed to d...

Страница 34: ...function sec 1 50 Interval function sec 1 50 Precise regulation of pressure and time in steps of 1 5 or 10 Setting of individual pressure in each chamber included Number of independent channels 1 Adj...

Страница 35: ...35 The basic Carepump set includes Central unit device Power cord 7 Contents of the package A B A B...

Страница 36: ...th air duct hip and abdomen cuff with air duct short pants cuff with air duct safety button with a cable for the patient allowing him to stop the treatment at any time air duct splitter for simultaneo...

Страница 37: ...ing it in a manner inconsistent with the manual may result in the loss of warranty and in case of the need for intervention by an authorized service may result in treating it as a com mercial service...

Страница 38: ...her accessories contact an author ized service center or the manufacturer The touch screen works incorrectly or it does not work at all Disconnect the device from the power supply Clean the front pane...

Страница 39: ...wer input on the back of the case Below the socket for the power cord there is a fuse drawer A Open it and pull out the glass fuses placed inside If the plate inside the glass fuse is intact it means...

Страница 40: ...referably every 12 months If you experience any failure with your machine or its ac cessories or if you notice a device malfunction contact the distributor immediately In case of any disturbance in th...

Страница 41: ...e carried out by an au thorized service center with personnel trained by the man ufacturer In case of problems with locating the service please contact the device distributor Only an authorized servic...

Страница 42: ...gation as indicated by the following symbol on the label of the device The product should be disposed of in dedicated places and points for electronic waste and in accordance with the pre vailing law...

Страница 43: ...tions are met Attention 1 Before sending the machine to service contact the dis tributor and remember to include the technical passport of the device inside the package 2 Pack the device in accordance...

Страница 44: ...Lite 4 Carepump Lite 6 are in conformity with the provisions of the following EC directives EC Electromagnetic Compatibility EMC Directive 2014 30 EU EC Low Voltage Directive LVD 2014 35 EU and that...

Страница 45: ...Manufacturer BardoMed Sp z o o ul Konecznego 6 66 31 216 Krak w NIP 945217693...

Отзывы: