1. Introduction 

 

CuroCell S.A.M.® is a constant low pressure reactive air mattress system that can be used in all 

kinds of caring environments and it is very functional and cost effective.  

 

     Always read the instructions for use prior to use.

 

 

1.1 General information

 

This system is a medical device with CE marking in accordance with Medical Device Regulation 

MDR (2017/745). 
The system has been fully or partially designed and verified in accordance with the standards 

listed in section 8.2 Standards. 
According to statutory regulations made by the authorities regarding medical devices, the 

manufacturer is required to report all accidents or incidents involving the products. We would 

be very grateful for all information involving accidents or incidents relating to our products, by 

reporting them immediately to us, at Care of Sweden. 

 

1.2 Intended use

 

The system consists of a pump and a mattress and are used for prevention and as an aid in the 

treatment of pressure ulcers/pressure injuries (PU/PI). The system can also be used in pain 

therapy. 

 

1.3 Indications

 

For users with pressure ulcers/pressure injuries (PU/PI) up to and including category 2 ¹. As 

well as for users in risk of PU/PI.  

 

1.4 Contraindications 

There are no known contraindications.

 

 

1.5 User application   

The systems can be used in all kinds of health care environments. 

It is available in three 

versions.

 CuroCell S.A.M.® and CuroCell S.A.M.® CF10

 are 

overlay

 

mattresses and are placed 

on an underlying mattress. 

CuroCell S.A.M.® CF16

 is a replacement mattress placed directly on 

the bed base. The 

recommended user weight is up to 200 kg.

 

 

 

 

 

 

 

 

 

 

 

Содержание CuroCell S.A.M.

Страница 1: ...Instructions for use 2020 12 08 CuroCell S A M Constant low pressure reactive air mattress system Instructions for use item number 95 001298...

Страница 2: ...2 Product description 5 3 Installation and start up 7 4 Reuse cleaning and reconditioning 9 5 Storage 11 6 Maintenance 11 7 Troubleshooting 11 8 Technical description 12 9 Spare parts 14 10 Options an...

Страница 3: ...OTE Be particularly aware of trapping damage when using bed rails 5 Never use the product if the pump power cable or plug is defective the pump casing is damaged or it is not functioning properly Cont...

Страница 4: ...ful for all information involving accidents or incidents relating to our products by reporting them immediately to us at Care of Sweden 1 2 Intended use The system consists of a pump and a mattress an...

Страница 5: ...can be found on our website www careofsweden com or you can order a copy from our Customer services by sending your details to export careofsweden se 2 Product description 2 1 Pump 1 Front panel 2 Tub...

Страница 6: ...air supply and pressure in the mattress increase When the knob is turned counter clockwise the air supply and pressure in the mattress decrease Appropriate air pressure is set between 1 and 8 The tabl...

Страница 7: ...e bed or place it on a flat sturdy surface Place the power cable so there is no risk of stumbling over it run it over with the bed wheels or getting it jammed when raising or lowering the bed NOTE A p...

Страница 8: ...nderlying mattress There should be about 1 2 fingers margin If you can feel the user s body resting on your hand you need to increase the comfort settings to next highest step Wait 10 minutes and perf...

Страница 9: ...pose such as alcohol with or without tenside surfactants oxidizing solutions chlorine max 1 or hydrogen peroxide max 1 5 If using a different agent choose one that does not affect the pump MATTRESS CO...

Страница 10: ...utions chlorine max 1 or hydrogen peroxide max 1 5 Let it dry 5 Wipe off the work surfaces using a suitable detergent and or disinfectant 6 Put the mattress together If cells for any reason have be co...

Страница 11: ...ectly Pressure might be incorrectly set Adjust it according to the on the bed weight table and wait a few minutes Check again using the hand check see chapter 3 4 Check that the CPR valve is closed Th...

Страница 12: ...0x200 210x10 16 cm Weight 2 7 kg NOTE Care of Sweden reserves the right to modify the product specification at any time 8 2 Standarder Systemet har testats och godk nts enligt f ljande standarder r ti...

Страница 13: ...70 C CE marked acc to Medical Device Regulation MDR 2017 745 Machine wash at 95 C Directly on bed base Tumble dry Place on top of existing mattress Do not iron Recommended user weight User informatio...

Страница 14: ...31 105200 452 105200 CC 905616 105x210 cm 431 105210 452 105210 CC 905617 120x200 cm 431 120200 452 120200 CC 905618 120x210 cm 431 120210 452 120210 CC 905620 Air cells incl tubes with CPR Air cells...

Страница 15: ...28047 90x210 cm CC 928067 90x210 cm CC 928048 105x200 cm CC 928068 105x200 cm CC 928050 105x210 cm CC 928070 105x210 cm CC 928049 120x200 cm CC 928069 120x200 cm CC 928051 120x210 cm CC 928071 120x210...

Страница 16: ...and lasts longer During the expected lifespan of the product some parts of the pump will experience wear and will require servicing by a certified service technician in order to retain good functional...

Страница 17: ...recycling centre and sorted as electronic waste Mattress A used CuroCell mattress must be taken to a recycling centre and sorted as combustible waste For more information contact Care of Sweden or you...

Страница 18: ...18 13 Own notes...

Страница 19: ...19...

Страница 20: ...et www careofsweden com Address Care of Sweden AB P O Box 146 SE 514 23 Tranemo SWEDEN Visit address Fabriksgatan 5A SE 514 33 Tranemo SWEDEN Cargo address Byns v g 4A SE 514 33 Tranemo SWEDEN Distrib...

Отзывы: