POWER SUPPLY

1. Plug power cord into a fully grounded, 3-pronged wall receptacle that
supplies 115 VAC power. If an extension cord must be used, it should be a 3-
wire, fully grounded type.

2. If a 2-pronged wall receptacle is encountered, it is the responsibility of the
customer to contact a qualified electrician to replace the 2-pronged
receptacle with a properly grounded 3-pronged wall receptacle or have a
grounding adapter properly installed in the 2-pronged receptacle.

To avoid electrical hazard, DO NOT, under any conditions,
cut, remove, or bypass the power cord grounding prong!

3. On Models requiring 230 VAC power, it is the responsibility of the
customer to have a qualified electrician install the proper power cord plug
which conforms to national electrical codes and local codes and ordinances.

START-UP

1. Switch power ON. Power switch is located on right end of backside of
scale.

2. When switched on, scale will perform a brief test routine shown on all
displays. When scale is ready for use, the display will show zero.

KEYBOARD FUNCTIONS

lb/kg*

lb/kg key allows scale to measure in pounds/ounces or
kilograms/ grams. Pressing lb/kg key after scale locks
weight on display converts lb/oz to kg/g and conversely.
Appropriate indicators illuminate.

TARE/ZERO

TARE/ZERO subtracts tare weight, zeros scale, and
displays NET weight. With scale unloaded, press
TARE/ZERO key to reset scale to zero.

CLEAR WT.

CLEAR WT. key clears locked weight from display. If
infant movement causes display to continuously flash
during weighing operation, press CLEAR WT. to lock
weight on display

* Does not apply to Models 6732 and 6733.

Содержание Detecto 6732

Страница 1: ...Neonatal Scale 6732 6733 6735 and 6736 Operation Manual Cardinal Scale Mfg Co P O Box 151 Webb City MO 64870 Ph 417 673 4631 Fax 417 673 5001 Printed in USA web site www detectoscale com 8525 M039 O1 Rev E 10 03 ...

Страница 2: ...frequency energy and if not installed and used in accordance with the instructions manual may cause interference to radio communications It has been tested and found to comply with the limits for a Class A computing device pursuant to Subpart J of Part 15 of FCC Rules which are designed to provide reasonable protection against such interference when operated in a commercial environment Operation o...

Страница 3: ...onic circuitry permits accurate weighing of a moving infant and locks weight on the display even after removal of the infant until CLEAR WT key is pressed This manual will help to acquaint you with the proper setup operation and care of the Model 6700 Series Neonatal Scale Please read it before attempting to use the scale and keep it handy for future reference FEATURES AND SPECIFICATIONS Easy to r...

Страница 4: ...2 COMPONENTS AND CONTROLS ...

Страница 5: ... unit or proper operation of the scale may be inhibited 3 Before mounting infant tray onto scale if necessary level scale by adjusting the four legs on bottom of scale up or down until the bubble in the level gage located on the weigh bridge is centered as shown in illustration below After level is achieved lock legs in place by tightening lock nuts on legs against scale bottom Adjusting Legs Leve...

Страница 6: ... proper power cord plug which conforms to national electrical codes and local codes and ordinances START UP 1 Switch power ON Power switch is located on right end of backside of scale 2 When switched on scale will perform a brief test routine shown on all displays When scale is ready for use the display will show zero KEYBOARD FUNCTIONS lb kg lb kg key allows scale to measure in pounds ounces or k...

Страница 7: ...5 OPERATION ...

Страница 8: ... sudden shocks 2 DO NOT press the keys too hard Only a light touch is necessary Keys must be activated only by your finger DO NOT use a sharp object 3 DO NOT submerge the scale in water or spray water directly on the scale Clean the cover and infant tray with a damp cloth and mild detergent DO NOT use abrasives 4 DO NOT use acetone or other volatile solvents for cleaning 5 DO NOT autoclave the inf...

Страница 9: ...ded Lo Indicates gross weight is less than 0 BEFORE YOU CALL FOR SERVICE Customer satisfaction is our prime objective If however any problems occur in connection with the operation of this scale please check the following points before calling for service 1 Is the power cord fully inserted into the wall receptacle 2 Is the power switch turned on 3 Is the wall receptacle receiving proper AC power 4...

Страница 10: ...346 BABY TRAY 3 427R1348 RULER FOR BABY SCALE 4 442R137 KEYPAD 6735 BABY SCALE 5 6540 1011 LEVELER SCREW 1 4 20x1 LONG S S 6 6610 4019 FUSE 1 4 ASB 3AG 1 4 ASB 7 6610 4046 FUSE HOLDER PANEL MOUNT KNOB TYPE 8 8535 D200 0A PCB ASSEMBLY 673X LED DISPLAY 9 8535 D201 0A PCB ASSEMBLY 6732 6735 CONTROLLER 3 2 5 8 4 1 9 6 7 ...

Страница 11: ...2 ...

Страница 12: ...U N A M E D S I G U N E D F A C T R E D I N U S A U N A M E D S I G U N E D F A C T R E D I N U S A ...

Отзывы: