Figure 2
Clamp
Battery Charging Port
Therapy Timer / Display
Additional Battery Information
NOTE:
The Cardinal Health™ SVED® is designed to permit use of the product while the internal battery is charging.
The device will continue to operate properly while charging is taking place.
Average Battery Life
The specified battery life of the Cardinal Health™ SVED® device with a fully-charged battery is approximately 18 hours.
The actual life is dependent on the integrity of the dressing. A leak in the dressing can reduce overall battery
longevity significantly.
Average Time for Recharging
To ensure the battery has been fully charged, the device should be connected to an A.C. supply for approximately
3 hours. After approximately 2 hours of charging, the device will have achieved 80% of total battery capacity.
Low Battery Alarm
While running on battery, a low-battery alarm will “chirp” every 10 seconds and the OFF button will begin flashing
when remaining capacity of the battery is less than 20%. Typically, the device will continue to operate for approximately
1 hour after the low-battery alarm is activated.
Low Battery Shutoff
If the battery charge falls below a critical level, the device will shutoff automatically and therapy will be discontinued.
At this point, the device must be plugged into an A.C. power source for therapy to resume. Once the A.C. Adapter is
plugged in, pressing the ON button will restart the device.