Recharging the Battery
Plug the power cord from the A.C. Adapter into the power connector on the side of the device. Plug the A.C. Adapter into
a suitable 120 VAC, 60 Hz wall outlet.
When the device is connected to an AC power source, the GREEN LED below the power connector will illuminate
indicating AC power is present and the AMBER LED will illuminate when the battery is charging.
Once the battery is fully charged, the AMBER LED will extinguish indicating the charge cycle is complete.
When the Cardinal Health™ SVED® device is disconnected from the AC power source, the device will automatically
switch over to the internal battery and continue to operate without interruption.
2.3 Changing the Canister
If the canister fills with fluid, an alarm will sound to let you know you must change the canister. Refer to the Alarm
Troubleshooting guide found in the
3.3 ALARM OPERATION
section of this manual.
The canister full alarm begins when the canister is 90% full, but the device will continue to work until the canister
completely fills.
1. Clamp the tubing closed
(Figure 3).
2. Turn the device OFF by pressing the button.
3. Remove the blue SpeedConnect™ tubing from the top of the canister
(Figure 3)
. Twisting the tapered connector
will make removing the Suction tube from the canister easier.
Figure 3
SpeedConnect™ Tubing Connector
Clamp
Canister