Patient User Manual

Cat. 

6701132

Cardinal Health™ SVED®

Содержание SVED 6701132

Страница 1: ...Patient User Manual Cat 6701132 Cardinal Health SVED ...

Страница 2: ...Patient User Manual ...

Страница 3: ...lugged in whenever possible to keep the battery fully charged Always take the power cord with you when you leave home Keep the Cardinal Health SVED device upright to avoid a false canister full alarm Keep the Cardinal Health SVED device turned on at all times unless there is bleeding from the wound or instructed by your Healthcare Professional DonotchangethesettingsontheCardinalHealthSVEDdeviceunl...

Страница 4: ...iedoutonlybyqualifiedpersonnelauthorizedbyCardinalHealth The electrical installation of the room in which the device will be used complies with the appropriate national electrical standards The product must be used in accordance with this manual and all associated labeling and the Instructions for Use Any device that does not function as expected must be returned to Cardinal Health Notice to Users...

Страница 5: ...ice 6 2 2 Charging the Battery 6 2 3 Changing the Canister 8 2 4 Disconnecting the Device 10 3 Operating Instructions 11 3 1 Power ON OFF 11 3 2 Power up Procedure 11 3 3 Alarm Operation 12 4 Care and Cleaning 13 4 1 Disposal of Dressings Canister and Other Disposables 13 4 2 Cleaning the Device 13 4 3 AC Power Adapter Inspection 14 5 Specifications 15 5 1 Symbols 16 5 2 Electromagnetic Compatibil...

Страница 6: ...ounds active bleeding difficult wound hemostasis or who are on anticoagulants When placing the Cardinal Health NPWT foam dressing in close proximity to blood vessels or organs take care to ensure that they are adequately protected with overlying fascia tissue or other protective barriers Exposed tendon nerves or blood vessels should be protected by moving available muscle or fascia over them or by...

Страница 7: ...e serious With or without the Cardinal Health SVED device infection can lead to many adverse complications including pain discomfort fever gangrene toxic shock septic shock and various other complications With signs of more serious complications of infection discontinue the use of the Cardinal Health SVED device until the serious infection is diagnosed and properly treated CARDINAL HEALTH NPWT DRE...

Страница 8: ... 2 2 Charging the Battery The Cardinal Health SVED device has an internal battery that provides up to 18 hours of operation from a single full charge When the battery is running low an alarm will sound to let you know you must plug in the device to charge the battery See section 3 3 ALARM OPERATION of this manual for AlarmTroubleshooting 1 Plug the device s A C Adapter into a suitable wall outlet ...

Страница 9: ... battery has been fully charged the device should be connected to an A C supply for approximately 3 hours After approximately 2 hours of charging the device will have achieved 80 of total battery capacity Low Battery Alarm While running on battery a low battery alarm will chirp every 10 seconds and the OFF button will begin flashing when remaining capacity of the battery is less than 20 Typically ...

Страница 10: ...r source the device will automatically switch over to the internal battery and continue to operate without interruption 2 3 Changing the Canister If the canister fills with fluid an alarm will sound to let you know you must change the canister Refer to the Alarm Troubleshooting guide found in the 3 3 ALARM OPERATION section of this manual The canister full alarm begins when the canister is 90 full...

Страница 11: ...ur Healthcare Professional how to properly dispose of a used canister 6 To install a new canister hold the new unused canister at the bottom and slide upwards into the holder 7 Align the two short ports and press upwards until it clicks into place 8 Reconnect the SpeedConnect tubing to the canister Gently twist and push the blue tapered connector on the open port of the canister just enough to sec...

Страница 12: ...d at least 22 hours out of every 24 hour period ContactyourHealthcareProfessionalimmediatelyiftherapyisterminatedorisoffformorethan2hoursina24hourperiod Your dressing may need to be changed 1 Clamp the tubing closed Figure 5 2 Turn the device OFF by pressing the button 3 Remove the blue SpeedConnect tubing from the top of the canister Figure 5 Twisting the tapered connector will make removing the ...

Страница 13: ... located on the front top of the control panel and control the application of power to the therapy device 3 2 Power Up Procedure When turning ON the Cardinal Health SVED device this is what you should expect to happen 1 Press the ON button All LED indicators will illuminate for 1 second during the power on self test after which the ON button will steadily light green 2 Dressing should collapse ind...

Страница 14: ...les in the drape They can occur if the drape snags on clothes or bed sheets There is an air leak in either the dressing or the tubing connections Clamp the tubing Turn device off by pressing the OFF button Press the canister release button above the canister and slide the full canister out Cap and dispose of properly Slide new canister in align the short ports and click into place Open the clamp a...

Страница 15: ...otential hazard to the patient or Caregiver WARNING The Cardinal Health SVED device should only be used with the supplied A C Adapter Use of an incorrectly rated adapter could create a shock hazard for the patient or caregiver and or severely damage the device 4 1 Disposal of Dressings Canister and Other Disposables Your Healthcare Professional should remove your dressings tubings clamps used cani...

Страница 16: ...th SVED device should only be used with the supplied A C Adapter Use of an incorrectly rated adapter could create a shock hazard for the patient or caregiver and or severely damage the device WARNING Avoidspillingliquidonanypartofthetherapydevice Liquidscancausecorrosionwhenleftonelectronic controlswhichcanleadtofailure Componentfailuremaycausethetherapydevicetooperateerratically possiblycausinga ...

Страница 17: ...ered Equipment IPXO Battery Duration Fully Charged up to 18 hours Electrical External Power Supply Input 100 240 VAC 50 60Hz 700 mA External Power Supply Output 15 VDC 2 Amps Patient Enclosure leakage Current 100 Micro amps Environmental Conditions Storage Conditions Temperature Range 10 F 12 C to 110 F 43 C Relative Humidity Range 20 95 Non condensing Atmospheric Pressure Range 50 kPa to 110 kPa ...

Страница 18: ...ful effects of water IPXO Alternating Current Device Consult Instructions for Use Expiry Date Date of Manufacture Lot Batch Number Fragile Keep Dry Single Use Only Method of Sterilization Ethylene Oxide Storage Conditions Conforms with the Medical Device Directive 93 42 EEC and has been subject to the conformity procedures laid down in the council directive Dressings ...

Страница 19: ...e Cardinal Health SVED Device Separation distance according to frequency of transmitter in meter s 0 01 0 1 1 10 100 0 12 0 38 1 2 3 8 12 0 12 0 38 1 2 3 8 12 0 23 0 73 2 3 7 3 23 The Cardinal Health SVED device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Cardinal Health SVED device can help prevent electrom...

Страница 20: ...mpliance Level 3V rms 3V m 3V rms 150 kHz 80 MHz 3V m 800 MHz 2 5 GHz Portable and mobile RF communications equipment should be used no closer to any part of the Cardinal Health SVED device including cables than the recommended separation distance calculated from the equation appropriate to the frequency of the transmitter Recommend separation distance d 1 2 P d 1 2 P 80 MHz to 800 MHz d 2 3 P 800...

Страница 21: ...e 47 4500 Accessories Cardinal Health SVED I V Pole Bed Hanger Adapter 47 5500 Cardinal Health SVED Canvas Carry Bag 47 9500 Cardinal Health SpeedConnect Tubing 47 2000 Cardinal Health NPWT Y Connector 47 2500 Cardinal Health NPWT Polyurethane Drape 10 per pkg 47 7000 Cardinal Health NPWT SensiSkin Drape 10 per pkg 47 7100 Cardinal Health NPWT Irrigation Tubing with SpeedConnect 47 6000 NOTE Part ...

Страница 22: ...th Waukegan IL 60085 www cardinalhealth com CustomerServiceNPWT cardinalhealth com ForquestionsorcommentsregardingthecontentofthisUser sManual pleasecontactCardinalHealthattheabove address PleasecontactCardinalHealthCustomerServiceat1 866 484 6798forissuesconcerningtheproductanditsuse Alwaysconsultaphysicianandproductinstructionsforusepriortoapplication Caution Federal law restricts these devices ...

Страница 23: ...21 Patient User Manual ...

Страница 24: ...h com 2015 Cardinal Health All Rights Reserved CARDINAL HEALTH the Cardinal Health LOGO SVED SENSISKIN and SPEEDCONNECT are trademarks or registered trademarks of Cardinal Health US Pat 7 532 953 7 608 066 8 066 243 EMERGO EUROPE Molenstraat 15 2513 BH The Hague The Netherlands ...