19
Clinician U
ser Manual
4.6 Battery Operation
NOTE:
The Cardinal Health™ NPWT PRO family of devices is designed to permit use of the product while the internal
battery is charging. The therapy device will continue to operate properly while the battery is charging.
Battery Life
The specified battery life of the Cardinal Health™ NPWT PRO family of devices with a fully-charged battery and a well-sealed
dressing is up to 24 hours. The actual life is dependent on the integrity of the dressing. A leak in the dressing can reduce
overall battery longevity significantly.
Average Time for Recharging
To ensure the battery has been fully charged, the device should be connected to an A.C. supply for approximately 3 hours.
After approximately 2 hours of charging, the device will have achieved 80% of total battery capacity.
Low Battery Alarm
While running on battery, a low-battery alarm will activate when remaining capacity of the battery is less than 20%
(See
4.7 ALARM OPERATION
). Typically, the device will continue to operate between 30 minutes and 1 hour after the
low-battery alarm is activated.
Low Battery Shutoff
If the battery charge falls below a critical level, the device will shutoff automatically and therapy will be discontinued.
At this point, the device must be plugged into an A.C. power source for therapy to resume. Once the A.C. Adapter is
plugged in, pressing the ON
button will restart the device.
Recharging the Battery
Plug the power cord from the A.C. Adapter into the Battery Charging Port on the side of the therapy device. Plug the A.C.
Adapter into a suitable 120 VAC, 60 Hz wall outlet.
When the device is connected to an AC power source, the green “power” LED on the front of the device will illuminate
indicating AC power is present and the amber “charging” LED, located just below the “power” LED, will illuminate when
the battery is charging.
Once the battery is fully charged, the amber LED will extinguish indicating the charge cycle is complete.
When the Cardinal Health™ NPWT PRO family of devices is disconnected from the AC power source, the devices will
automatically switch over to the internal battery and continue to operate without interruption.