Screen/Button
Display
Function
Summary
1
Initial Screen
SEr 05
Temp of unit
Allows the user to see what temperature the unit is
designed to run.
Varies
CPU Checksum
Shows the current checksum of the CPU board for
validation.
Varies
Relay Board Checksum
Shows the current checksum of the Relay board for
validation.
tyP
Model Type
Unit type 05 for +5 refrigerator. This parameter should
never be changed.
2
Press Scan (
SCAN
)
CFg
Type of unit
Type of unit (01). This parameter should never be
changed.
3
Press Scan (
SCAN
)
CuFt
Size of the unit
Setting for size of the unit.
4
Press Scan (
SCAN
)
Pd oFSt
Control Probe Pulldown
Offset
The difference in temperature between the control
probe value and the average compartment
temperature that determines when the control system
changes from pull down mode to steady state operation.
This can be useful to adjust if the average compartment
temperature has an undershoot or overshoot relative
to set point during pull down mode. This offset is also
used on some refrigerator models as a response to door
openings.
5
Press Scan (
SCAN
)
Cnt oFSt
Control Probe Offset
The difference in temperature between the control
probe value and the average compartment
temperature that is used to control the cooling system
to the unit set point during steady state operation. This
can be useful to adjust if there is a discrepancy between
the unit set point and the average cabinet temperature.
6
Press Scan (
SCAN
)
diS oFSt
Display Probe Offset
The difference in temperature between the display
probe value seen on the User Interface display and the
average compartment temperature at the unit set point
during steady state operation. This can be useful to
adjust the display during unit calibration procedures.
7
Press Scan (
SCAN
)
Cnt ucl
Control Upper Hysteresis
The degree increase in temperature from the set point
that will trigger the unit to begin cooling to prevent the
unit from getting too warm.
8
Press Scan (
SCAN
)
Cnt lcl
Control Lower Hysteresis
The degree decrease in temperature from the set point
that will trigger the unit to stop cooling to prevent the
unit from getting too cold.
9
Press Scan (
SCAN
)
dEF Int
Defrost Interval
The time interval between defrost cycles in hours.
Table 8. Service Parameters