1000DF00005 Iss. 2
20/44
Basic Features (Continued)
Pressure Level
1. To check and adjust the pressure level press the
e
button. A bar graph will be displayed showing the pressure alarm level and the
current pressure level.
2. Press the
f
keys to increase or decrease the alarm level. The new level will be indicated on the display.
3. Press
OK
to exit the screen.
A
During PURGE, BOLUS and INDUCTION the pressure limit alarms are temporarily increased to their maximum level.
For TCI operation a threshold rate may be set above which the pressure limit alarms are temporarily increased to
their maximum level.
Note: This is not applicable in TCI mode.
If Rate Titration is
enabled
the rate can be adjusted
while infusing
:
1. Select the new rate using the
f
keys.
The message < START TO CONFIRM > will flash on screen and pump continues to infuse at the original rate.
2. Press the
b
button
to confirm the new infusion rate and start infusing at the new rate.
If Rate Titration is
disabled
the rate can only be adjusted
whilst on hold:
1. Press the
h
button to put the pump on hold.
2. Select the new rate using the
f
keys.
3. Press the
b
button
to start infusing at the new rate.
Rate Titration
Note: This
only
applies to TCI mode.
Concentration Target Titration allows the rate to be adjusted while infusing:
1. Select the new target using the
f
keys.
The pump status is shown as TITRATE and the pump continues to infuse at the original concentration target.
2. Press
the
b
button to confirm the new concentration target and start infusing at the new rate. If the new concentration target
setting exceeds or is under a Soft Alert, confirmation is required before infusion can resume.
Concentration Target Titration
Clear Volume
Note: Clear Volume is not permitted in TCI mode or predictive TIVA mode.
This option enables the volume infused to be cleared.
1. Press the
VOLUME
softkey to display the
CLEAR VOLUME
option.
2. Press the
YES
softkey to clear the volume. Press the
NO
softkey to retain the volume.
Selecting YES resets the volume infused in the 24H LOG option.