Programming
4-13
Alaris
®
System Directions for Use
SpO
2
Modules Section
Channel Options
(Continued)
2.
To change
Saturation Averaging Time
, press applicable
soft key. Selectable options are
2
,
4
,
8
,
10
,
12
,
14
and
16
seconds.
•
FAST SAT
is enabled when
2
or
4
seconds is selected.
3.
To return
SPO2
Main
display, press
CONFIRM
soft key.
A
A
Increase
Decrease
Saturation Averaging
Time
4 Seconds
CONFIRM
>Press CONFIRM
with
Fast SAT
Changing Saturation Averaging Time
(Continued)
Changing Sensitivity Mode
1.
Press
Sensitivity Mode
soft key.
A
Pulse Beep Volume: Off
Sat. Averaging Time: 8
Limit Mode:
Adult
Channel Options
>Select an Option or
EXIT
EXIT
Sensitivit
y
Mode:
Normal
2.
To change
Sensitivity Mode
, press applicable soft key.
•
Normal
: Normal patient monitoring.
•
Maximum
: Improved low perfusion performance.
A
A
Normal
Maximum
Sensitivit
y
Mode
EXIT
>Select an Option or
EXIT
NOTE:
The sensitivity mode displays on the
SPO2
Main
display only
when
Maximum
is selected.
(8220)
(8220)
Содержание Alaris 8000 Series
Страница 8: ......
Страница 52: ......
Страница 54: ......
Страница 176: ......
Страница 242: ......
Страница 284: ......
Страница 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Страница 318: ......
Страница 338: ......
Страница 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Страница 340: ......
Страница 350: ......
Страница 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Страница 352: ......
Страница 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Страница 362: ......
Страница 377: ......
Страница 378: ......