MRF
16/22
MF 55-GB 1.21
31
Press scroll
3
Holdback
Value is displayed. Press up
S
or down
T
to set to ‘3’ degrees
Press scroll
3
Ramp Units is displayed. Press up
S
or down
T
to set to ‘Min’
Press scroll
3
Cycles is displayed. Default ‘1’
Press scroll
3
Segment 1 is displayed
Press scroll
3
Segment Type is displayed. Press up
S
or down
T
to select ‘Time’
Press scroll
3
Target SP is displayed. Press up
S
or down
T
to set to ‘1200’ degrees
Press scroll
3
Duration is displayed. Press up
S
or down
T
to set to ’2:25:00’.
Press scroll
3
Holdback Type is displayed. Leave at default setting ‘Off’.
Press scroll
3
Micro% is displayed. Leave this as it is.
Press scroll
3
Event outs is displayed. The first square will be flashing. Press up
S
or
down
T
to fill the first square in
.
Press scroll
3
The second square will flash. Leave this square blank
.
For segments 2,3,4 and 5 enter the following parameters and values:
Segment 2: Dwell, Duration 1:00:00, Micro % 100, Event outs
(Segment 3: Call, Call Prog 3, Call Cycles 1. Only on multi program controllers)
Segment 3(4): Step, target set point 800°C, Holdback High, Micro %100, Event outs
Finish the program with an End segment:
Segment 4(5): End, Dwell, Event outs
To run the program:
Press page
and scroll
3
at the same time return to the home page.
Press RUN/HOLD twice
To pause (hold) the program:
Press RUN/HOLD once
To reset the program:
Press RUN/HOLD once to pause (hold) it
Press RUN/HOLD for 3 seconds.