Calibration
First calibration is “0”. Perform a measurement on uncoated metal
for which you’re doing the calibration.
Wait for stable readings and put the base away from the device. If
needed, repeat the measurement.
Press and release “PWR” to save the result an go to the next
measuring point.
The screen will now display “F-1” and the device will also show
recommended calibration plate thickness for this point.
Perform the measurement on the base metal together with
recommended calibration plate on top.
Select (by pressing “MODE” and “UNIT”) the thickness of
measured calibration plate and press “PWR” to go to the next
calibration point.
Repeat the same procedure for each calibration point to follow.
Once the complete, the screen will display “8888” and the device
will go into “P” mode.
Repeat the whole process for the second base metal type.
Please note that if calibration was not performed properly for one
type of the base metal, the device will not be able to automatically
detect the base type.