Note: If you prefer to turn off the entire system every day, then allow at least 30 minutes
for system warm up before performing any calibration procedures or making any
measurements.
2.
Schedule for performing the Quality Assurance (QA) Tests
a. Setting Up QA Parameters --
Once, at initial system setup
(page 5-2)
b. Auto Calibration ----------------
Daily, before the first procedure
(page 5-5)
c. Constancy Test -----------------
Daily, before the first procedure
(page 5-10)
d. Chi-Square Test ----------------
Quarterly
(page 5-15)
e. MDA Test ------------------------
Quarterly
(page 5-19)
While performing any Quality Assurance test, ensure the following:
•
Consistent placement of the radioactive standard source in the detector is critical.
The source should not be leaning against the sides.
•
Keep background rates constant throughout the counting process by ensuring that no
radioactive sources or other dosed patients are in the immediate area.
Note: Perform all the needed QA procedures for one detector and then, repeat the
same for the second detector, if applicable.
3.
Record Keeping
– After the system is setup and all the QA Tests have been completed
successfully, it is recommended that a list be printed of the system parameters for your
records (Refer to the
section on page 5-22). These
are all the default values used while doing various measurements (Thyroid Uptake,
Bioassay, Wipe Test etc.). In case of suspicious results in any module, confirm the Setup
values in that module with the values in your records. All QA results are stored
automatically and may be accessed at any time (Refer to the
section on page 5-24).
SETTING UP QA PARAMETERS
To access the QA tests, click the
button or press
Alt+Q
on the CAPTUS
®
3000 Main screen. Figure 5-1 Quality Assurance Main Screen appears as shown.
5-2
QUALITY ASSURANCE
July 14
Содержание CAPTUS 3000
Страница 2: ......
Страница 10: ...CAPINTEC INC CAPTUS 3000 INDEX WARRANTY DECLARATION OF CONFORMITY TOC 8 TABLE OF CONTENTS July 14...
Страница 14: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 22: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 86: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 124: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 266: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 284: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 292: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 304: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 334: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 336: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 337: ......
Страница 338: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...