If the Chi-Square value is out of range, check the following:
•
The positioning of the rod source, especially for the Probe detector.
•
Ensure that there are no other sources (including dosed patients) in the area.
If a problem is found with any of the two items above, then correct the problem and
repeat the Chi-Square Test.
In no problems are found with the above items, the out of range Chi-Square value may
indicate a problem with the counter/timer mechanism of the system or the detector itself.
Contact Capintec’s only Authorized Service Center for service.
QUICK TIPS:
•
To print current results for record keeping, click the
button or press
Alt+P
on Constancy Test screen after the test is completed.
•
To view or print results for all the Quality Assurance tests performed to date,
click the
button or press
Alt+V
(Refer to the
section on page 5-24).
MDA TEST
CAUTION:
Calibration is recommended before making any measurements. Refer to
the
section on page 5-5.
This test measures counts and calculates activity in the ROI for the isotope selected in the
Setup. The measurement is performed with no isotope placed in the detector, which
therefore gives a value for the Minimum Detectable Activity (MDA).
1. Click the
button or press
Alt+M
on Figure 5-1 Quality Assurance Main
Screen. Figure 5-22 MDA Test Screen will appear.
July 14
QUALITY ASSURANCE
5-19
Содержание CAPTUS 3000
Страница 2: ......
Страница 10: ...CAPINTEC INC CAPTUS 3000 INDEX WARRANTY DECLARATION OF CONFORMITY TOC 8 TABLE OF CONTENTS July 14...
Страница 14: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 22: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 86: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 124: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 266: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 284: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 292: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 304: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 334: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 336: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 337: ......
Страница 338: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...