The following 

COMMAND

 table describes the available commands:  

Flash count

 

Command

 

Description

 

1 flash 

STATUS 

Shows the device STATUS. As answer the led perform a flash sequence as reported 

in the “STATUS” table. If the device is performing the TEST (

refer to TEST command

this command stops it.  

2 flashes 

TEST 

Enter in TEST mode and transmits status and measurements every 5 seconds. If 
the device is in STANDBY mode or it is out of radio range, this command forces the 
connection procedure to the WSN and the return to the operative mode. The TEST 
stops after 120 seconds. During TEST, the led continuously shows the STATUS to 
monitor the received radio signal quality. 

CAUTION: Measures acquired during 

TEST

 phase are NOT saved. 

3 flashes 

ENROLL 

Association  to  the  network:  must  be  used  when  the  device  has  not  yet  been 
included  in  a  network,  starts  the  entry  and  association  procedure  to  the 

basestation

 (

refer to “

WineCap System - User Manual R30

). 

4 flashes 
  
 
+ 4 flashes 

STANDBY 

Temporary device deactivation: the device is stopped. The sampling process and 
the radio are/is turned off losing the connection to the network. To reactivate, a 
TEST  command  is  necessary.  The  STANDBY  command  must  be  given  twice  to 
confirm it: at the first sequence the led flashes alternating RED and GREEN lights, 
waiting  for  the  second  confirm  sequence  within  15  seconds.  At  the  command 
execution the led flashes as the STANDBY status (

refer to 

WIRELESS MODE STATUS 

Table

). 

5 flashes 
 
 
+ 5 flashes 

FACTORY 

RESET 

The device performs the memory deleting procedure and goes in STOP status. All 
samples,  configuration  and  wireless  network  data  associated  are  LOST.  To 
reactivate the device a new association and configuration procedure is necessary 
(

ENROLL command

). Also in this case, the FACTORY RESET command must be given 

twice  to  confirm  it.  At  the  command  execution  the  led  flashes  as  the 

“PROBE/DATALOGGER  NOT  ASSOCIATED”  status 

refer  to  

WIRELESS  MODE  STATUS 

Table

). 

5 flashes 
 
 
+ 3 flashes 

LOGGER      

NO WSN 

As  the  previous  command  but  performs  only  the  WSN  deleting  procedure  and 
disassociate  from  the 

basestation

.  The  device  enters  in  LOGGER  STAND  ALONE 

mode: data are kept, and the sampling activity CONTINUES with previous setup. 
Command must be given with 2 sequences: 5 flashes and then 3 flashes. At the 
command  execution  wait  for  the  device  reboot.    At  the  STATUS  command, 

“LOGGER” will be the answer (

refer to 

STAND-ALONE MODE  STATUS Table

). A new 

association (

ENROLL command

) is possible to a new 

basestation

6 flashes 

 
+ 1 flash 

CH1 and CH2 

Reset  

The device carries out the procedure of zeroing the inclination channels of the X 
and  Y  axes,  determining  the  absolute  position  of  the  instrument  and  the 
consequent ROTATION OF THE AXES, aligning the internal references and saving 
the ZERO position. 

6 flashes 

 
+ 4 flashes 

CH4 

Reset 

The device carries out the zeroing procedure of the relative distance acquisition 
channel 

6 flashes 

 
+ 6 flashes 

Cancellation 

of resets 

The device performs the procedure for cancelling all the zeroing operations carried 
out on channels 1, 2 and 4. The initial values are restored. 

7 flashes 

Pointing 

The laser pointer is turned on. To interrupt this phase, it is necessary to issue the 
command again or wait for the next measurement. 

 

 

Picture 3 - Commands table

 

Содержание WineCap WSD15IIDIST

Страница 1: ...urement time For the average to be carried out correctly the Follower function on the pitch and roll channel active by default must be enabled Otherwise the average will be performed with the cadence of the Measurement time The setting of these times is configurable and modifiable through the WineCapManager software or through the cloud b Measure thresholds Two thresholds can be set for each slope...

Страница 2: ...nded for dynamic event displays By sudden event is meant as significant displacement related to the position of rest in which the instrument is working d Reset installation position After installation and fixing a 6 1 command can be issued see Picture 3 Commands table which resets the channels dedicated to the inclination The command starts a first acquisition which determines the absolute positio...

Страница 3: ...r pointer There is a laser aiming function by giving the command 7 see Picture 3 Commands table the laser is activated and allows a precise aim at the target This function ceases when a measurement is performed by the datalogger or if command 7 is given again It is recommended to use a white non glossy and possibly smooth material as a target 4 Thermometer operating logic The value reported on cha...

Страница 4: ...dify the device s operation mode The sampling interval must be set with the device in STAND ALONE refer to 7 mode and automatically the device s clock is aligned with the PC s clock to assure the temporal reference of the sample Sampling operations start may be selected disconnecting the USB cable or giving the proper command with the magnetic key refer to 10 Stand alone USB datalogger installatio...

Страница 5: ...TANDBY status refer to WIRELESS MODE STATUS Table 5 flashes 5 flashes FACTORY RESET The device performs the memory deleting procedure and goes in STOP status All samples configuration and wireless network data associated are LOST To reactivate the device a new association and configuration procedure is necessary ENROLL command Also in this case the FACTORY RESET command must be given twice to conf...

Страница 6: ...ifferent command is associated to confirm the command the WineCapKey must be removed from the sensible area immediately after the number of flashes corresponding at the desired command The TEST corresponds to the second pulse and activate the Field Measurer function The device will give back the radio signal quality through led flashes WIRELESS MODE STATUS Table Flash count Wireless mode Status Ra...

Страница 7: ...unt Stand alone mode STATUS 1 green flash 2 seconds long ACTIVE 2 red flashes 2 seconds long STANDBY Sequence of red flashes short 2 seconds long short FACTORY RESET INVALID datalogger clock PC connection required 11 Shutting off Reactivating the device If the device is shut off and left unused for a long time you can issue the STANDBY command refer to Picture 3 Commands table It corresponds to th...

Страница 8: ...Three axis accelerometer Inclination Measure range 6 5000 6 5000 Inclination Measure accuracy 0 5 of measure Inclination Measure resolution 0 0002 Temperature Transducer type NTC10KΩ Temperature Measure range 30 C 60 C Temperature Measure accuracy 0 5 C Range 30 C 0 C 0 2 C Range 0 C 60 C Temperature Measure resolution 0 01 C Relative distance Transducer type LASER distance meter Relative distance...

Страница 9: ...o cause unwanted behaviour Modifications or tampering not expressly approved by the Manufacturer could void the user s authorization to operate the equipment This equipment must be installed by qualified personnel This equipment contains a Class 2 laser The user should avoid pointing the laser directly into the eye while the device is powered The device has a weight of up to 1 6 kg Make sure that ...

Страница 10: ...2011 65 CE that restricts the use of substances in the manufacturing of electronic devices The WEEE logo on the label indicates that this product is compliant with the WEEE EC Directive This symbol valid only in the European Union countries indicates that the product it is applied to MUST NOT be discarded with ordinary household or industrial waste but must be sent to an authorized reception point...

Отзывы: