Caution!-This symbol alerts you to important operating
considerations or a potential operating condition that
could damage equipment. Refer to user's manual or
operation's manual for precautionary instructions.
I
CAUTION
II
CLASSIFICATION
III
External Equipment
Symbol Description
Class I :
No applied parts
Protection against harmful ingress of water is IPX0.
Not suitable for use in the presence of flammable anesthetic's or oxygen.
Mode of operation: Continuous.
External equipment intended for connection to signal input / output or other connectors, shall
comply with UL/EN 60601-1 for medical electrical equipment. In addition, all such combination
system shall comply with the standard IEC 60601-1-1, Safety requirements for medical electrical
systems. Equipment not complying with UL/EN/IEC 60601-1 shall be kept outside the patient
environment, as defined on the systems standard.
Safety Precautions
IV Intended Use
The equipment is intended to be used as a component of a medical patient monitoring system.
MEDICAL EQUIPMENT
WITH RESPECT TO
ELECTRIC SHOCK,
FIRE AND MECHANICAL HAZARDS ONLY IN
ACCORDANCE WITH UL60601-1 AND
CAN/CSA C22.2 NO.601.1
49CG