Construction of the Pneumatic Module
2.4
Replacing the filter
Loosen screws
Inlet valve
Pressure regulator
Pressure controller
Filter
Flow sensor
procedural method:
1. switch off the cabinet
2. remove the tubes (by pressing the unlocking ring towards the housing)
3. loosen the screws
4. replace the filter
Caution:
Observe the direction of installation (arrow)!
5. reconnect the tubes, tighten the screws
6. switch on the cabinet and check that the tubes are connected tightly