SERVICE MANUAL

Содержание ENDODRY

Страница 1: ...one click SERVICE MANUAL...

Страница 2: ...use of this publication is prohibited without the express written consent of Cantel Cantel reserves the right to make changes in the specifications shown herein without notice or obligation Contact yo...

Страница 3: ...ENDODRY 13 3 6 Changing air circulation filter 14 3 7 Test function of the fan 15 3 8 Check the Door s 15 3 9 Check Bar Code Reader 16 3 10 Check Touch Display 16 4 OPERATING SOFTWARE 17 4 1 License...

Страница 4: ...e work repairs and modifications carried out by unauthorized staff To be in possession of the technical documentation of the device does not provide any authorization by the manufacturer to technicall...

Страница 5: ...ry Check LAN cable and connection for damage Replace them if necessary 2 2 Opening of the control system To open the control system first open the front door after that you can loosen fixing screw Air...

Страница 6: ...ler Filter Flow sensor procedural method 1 switch off the cabinet 2 remove the tubes by pressing the unlocking ring towards the housing 3 loosen the screws 4 replace the filter Caution Observe the dir...

Страница 7: ...ndoSTORE neo discrepancy endoSTORE one click ENDODRYTM one click Drying phase 1 0 5 bar 20 min 0 5 bar 30 min Drying phase 2 0 3 bar 0 3 bar 60 min Fan phase 180 min 180 min Storage phase 0 1 bar 0 1...

Страница 8: ...s is displayed and indicates Press is displayed for ca 1 second Now key lock is activated current pressure is displayed 2 5 3 Pressure controller The pressure controller monitors the set minimum and m...

Страница 9: ...Press 4th time current pressure is displayed 2 5 3 3 Completing adjustment of the pressure regulator In case that the pressure regulator needs to be exchanged or restored to factory settings the follo...

Страница 10: ...times F3 adjustment of response time Press change between and Press until is displayed Confirm with and go beck to menu Press 2 times F7 adjustment of power saving mode Press change between and Press...

Страница 11: ...et Tests At the yearly maintenance the following functions need to be tested 3 1 Check the Drawer Function free movement in out distortion 3 2 Check the Tube Connections tightness free movement of the...

Страница 12: ...2 of 28 3 3 Check the air outlet of all nozzles not for one click models deformation mechanical influences spring function 3 4 Check the o rings of the nozzles damage pinch porosity missing o rings o...

Страница 13: ...e diagnostic menu For the test an air flow tester is put on to every nozzle while air pressure pin is pushed The float of the air flow tester needs to reach the red marked range completely To check th...

Страница 14: ...page 14 of 28 3 6 Changing air circulation filter Loosen screws of the cover Remove cover Change filter Change filter Fasten Cover...

Страница 15: ...ke filter of the fastening pin Loosen plug Pin and fasten on new fan 3 8 Check the Door s retighten all screws door handle locking tab check the locking unit retighten screws check the hinges retighte...

Страница 16: ...onnector or reader check function via Diagnostic Scanner if two sided version repeat 3 10 Check Touch Display pixel errors check touch functionality and calibrate if necessary Menu Standard Settings C...

Страница 17: ...a service technician should perform 4 1 License settings Login as admin Within license settings site 2 menu it is possible to activate further addons The license key is provided by the manufacturer 4...

Страница 18: ...s 1 bar Caution Change of process settings has a major influence on drying result It is highly recommended not to reduce the process settings If the customer wants a change in this setting this will h...

Страница 19: ...the endoscope again Caution A change of error air pressure malfunction relative could lead to displaying a successfully completed process even if endoscopes have had insufficient air pressure 4 4 Ser...

Страница 20: ...the installer file Endodry Endostore_setup3 X XXXX exe to one of the available USB ports on the cabinet 2 On the USB memory stick double click on the installer file Endodry Endostore_setup3 X XXXX exe...

Страница 21: ...page 21 of 28 4 Check the Generate unique computer name box Important This checkbox is deactivated by default On the first installation it must be set manually to be active...

Страница 22: ...ch need to be installed and press the Next button Each cabinet version can be installed as a Master or Slave device Th NDODRY e following cabinet versions including legacy cabinets are available E END...

Страница 23: ...page 23 of 28 6 Press the Install button to install the Endodry application...

Страница 24: ...page 24 of 28 7 Press the Finish button once the application indicates the installation is complete 8 Installation complete...

Страница 25: ...33 l min 5 Bar 7 SCFM 1 2 CFM 72 5 PSI 6 1 4 according to ISO 8573 1 2010 1 4 1 Dimensions cm Weight kg Operating conditions Storing conditions Maximum relative humidity 80 for temperatures up to 31 C...

Страница 26: ...page 26 of 28 6 Colour touch screen I IP 20 Display Electrical protection class IP code...

Страница 27: ...ly note the different value cabinet tests o function of the drawers drawer rails o tube connections o functionality condition of inlet air pins o o rings of the nozzles if applicable replace o air out...

Страница 28: ...page 28 of 28 7 Technical Service Information Date Inspection Signature...