Bath

Instruction 

& Operation 

Manual

Содержание TE-3000

Страница 1: ...TE 3000 Thermo Electric Temperature Bath Instruction Operation Manual ...

Страница 2: ......

Страница 3: ...gements 5 D Liquid and pneumatic connections 6 E Platform pump assembly 6 F Electrical connections 8 G Selecting a bath liquid 9 4 GENERAL OPERATIONS 11 A Filling the bath with liquid 11 B Application of power 12 1 Cold start 12 2 Warm start 13 C Self test sequence 13 D Normal display and keypad indications 15 E LED indicators 17 F Adjusting the liquid level in the bath 17 G Setting the target des...

Страница 4: ...D Commands queries responses and reports 26 6 MAINTENANCE 31 A Preventative scheduled maintenance 31 Cleaning of painted surfaces and front panel 31 Cleaning the fans and heat sinks 31 Cleaning the bath window 32 Draining the bath liquid 32 Cleaning the inside of the bath vessel 33 7 WARRANTY RETURN INFORMATION 35 Products limited warranty 35 Reagent and chemical warranty 35 Returning a product to...

Страница 5: ...agnostics and repair should be performed in conjunction with or solely by CANNON Instrument Company Please consult CANNON Instrument Company for further assistance beyond the content of this manual B Overview TE 3000 features The TE 3000 is a state of the art microprocessor controlled and thermoelectrically cooled constant tempera ture liquid bath It is totally self contained in two stackable benc...

Страница 6: ...ator running at a very high frequency Calibration The TE 3000 can be calibrated to any temperature reference standard for any desired Celsius or Fahrenheit temperature The controls for entering the target desired temperature and the calibration at this temperature are very simple and straight forward The instrument remembers these calibration corrections for every possible temperature as well as t...

Страница 7: ...he Upper Bath Vessel communication RS 232 and RS 485 connectors address selection switch download demand switch two air pump connections to the Upper Bath Vessel and the bath fluid pump switch reside on the rear panel of the Lower Control Chassis Figure 2 TE 3000 Front panel The front panel contains 1 DC power indicator lamp 2 two lighted power switches bath lighting 3 liquid crystal display 4 15 ...

Страница 8: ...apors from to the upper Bath Vessel to from the lower Control Chassis The sex and type of these connectors constrain their use in only one way the proper configuration Overflow jar 3 One glass jar with lid is provided to catch the bath overflow liquid and is to be mounted on the rear overflow jar support along with three pieces of small diameter silicon tubing and an in line pump with connector Fo...

Страница 9: ...nd to the CANNON Instrument Company immediately Damaged items Retain all packing materials until the instrument is connected and func tioning properly If any component s must be returned to CANNON Instrument Company the damaged item s should be pack aged in the original shipping container Refer to the final chapter of this manual for instructions on returning defective equipment Customers outside ...

Страница 10: ...rovide the bath agitation and subsequent stirring by recirculating the bath vapors Two silicon tube assemblies are provided which have plastic quick connect bayonet style connectors on their ends These tubes are connected from the air in out connections on the bottom Control Chassis to the air out in connections on the upper Bath Vessel Housing There are female connectors on both ends of one hose ...

Страница 11: ...eciprocating Pump and insert the pump end into the small hole on the Overflow Platform so the pump rests on the platform 3 Reattach the nipple adaptor removed in Step 2 this holds the unit in place 4 Attach the electric connector cable from the Reciprocating Pump to the receptacle marked PUMP on the Control Unit To connect the cable make sure to align it correctly with the Bath Unit female connect...

Страница 12: ...onnector carries the temperature sensor signal from the upper Bath Vessel Housing down to the lower Control Chassis labeled PROBE This connec tion is keyed and it operates as a bayonet with a simple push in no twisting The red dot on the plug must be facing upward or inser tion will not be permitted When this connector is properly inserted a small click can be felt Removal is accomplished with a s...

Страница 13: ...nd the liquids suitable for use in those ranges S D I U Q I L H T A B 0 0 0 3 E T E G N A R E R U T A R E P M E T S D I U Q I L H T A B E L B A T I U S C 0 1 o t C 0 3 l o h o c l A l y h t e M C 0 2 o t C 0 1 l o h o c l A l y p o r p o s I l o h o c l A l y h t E C 0 3 o t C 5 s l i O y t i s o c s i V w o L r e t a W CAUTION Methyl alcohol methanol is very close to the ideal liquid it can be us...

Страница 14: ...a r e p m e T r e h t o t s o M s d i u l f c i n a g r o t s o m f o y t i c a p a c t a e h e h t e c i w t t u o b a s a h r e t a W r e t a w f o y t i c a p a c t a e h e h t f o f l a h t u o b a e v a h l l i w s d i u l f h t a b r o f s e c i o h c t a n o i t a r e p o r o f d i u l f h t a b a s a r e t a w e s u o t t p m e t t a T O N o D r e w o l r o C 2 f o s e r u t a r e p m e t ...

Страница 15: ...rflow liquid will spill down the overflow tube into the jar Bath volume changes The amount of extra liquid available in the overflow jar will depend upon what temperatures you will be operating the bath Also some bath liquid will be displaced by the addition of instruments and thermometers into the bath and thus cause the level to rise The bath liquid will expand if warmer than room temperatures a...

Страница 16: ...ector which plugs into the power inlet receptacle on the rear of the lower Control Chassis Be CERTAIN that you are plug ging into the correct mains voltage as defined on the rear panel identifi cation label Both front panel power switches should be OFF while this power cord is being installed Turning on the TE 3000 Turn on the main BATH power switch The LIGHTS switch is operable only if the BATH p...

Страница 17: ...ting several seconds turn the BATH power switch to the ON position and wait for the display to indicate either a warm or cold start again If the bath comes up in a cold start condition and starts its normal self test sequence it is possible that a momentary power interrup tion caused the warm start condition and no further action is required If the bath continues to enter the warm start condition ...

Страница 18: ... reset the Bath If any test fails during this power up self test sequence it will be noted on the display for that particular test The displays will resemble the previous sequence shown above but if any failure is encountered the OK will be replaced by BAD and PASSED will be replaced by FAILED All test sequences will be activated however at the conclusion of the test sequence the ENTER key will fl...

Страница 19: ...rget temperature is much colder than the actual temperature of the bath so the display indicates that it is cooling The COOLING indicator green LED on the right side of the front panel should be fully illuminated Heating indication In the following example the target temperature is much warmer than the actual temperature of the bath so the display indicates that it heating The HEATING indicator re...

Страница 20: ... C 1 1 Keypad The keypad on the front panel of the TE 3000 see Figure 8 consists of fifteen keys white numbers 0 through 9 a white combination minus sign and decimal point key a yellow SET TEMP erature key a yellow CAL ibrate TEMP erature key a yellow MENU key and an orange ENTER key The bottom row of five keys non numerical keys have the capability of being illuminated to assist the operator by i...

Страница 21: ... the liquid level in the bath Assuming that the bath has already been filled according to the instruc tions as provided in Chapter 4 A the user must adjust the level of the bath liquid as the bath temperature is decreased Since the bath liquid decreases in volume as the temperature decreases the operator must add liquid to the bath vessel to maintain proper circulation Maintaining bath liquid leve...

Страница 22: ...ange of the bath Incorrect settings If a temperature is entered which is outside of the operational tempera ture range of the TE 3000 the following display will be shown on the LCD and the ENTER key will flash RANGE ERROR Range is 43 to 33 C or 45 to 91 F ENTER to continue High temp adjustments If the bath was previously operating at a temperature below 20 Celsius and a new temperature is entered ...

Страница 23: ...d by the bath in this memory and will be used whenever that particular target temperature is entered Fractional degrees will use the closest whole number bin In the following example the target temperature is 20 degrees Celsius and the bath temperature is stable and controlling well within 01 degrees When the CAL TEMP key is depressed the display will show Enter Actual Bath Temp for Bin Offset at ...

Страница 24: ...nstrument Company for further assistance CALIBRATION ERROR Calib amount is too large Can t store ENTER to continue I Menu selections Whenever the bath is in the normal or quiescent mode the MENU key may be depressed The MAIN MENU will be entered and the following display will be shown MAIN MENU 1 Change to F 2 Bath Temp Offsets 3 Comm Setup _ From this MAIN MENU the user may enter any number from ...

Страница 25: ...gnificant differences become apparent between the user supplied reference tem perature and the displayed temperature of the bath or if CAL TEMP entries were made when the bath was not stable Once this clear command is issued new calibration entries may be made in the CAL TEMP menu ABOUT TO CLEAR ALL CALIB CONSTANTS ARE YOU SURE ENTER Yes MENU No b Setting the bath general offset Whenever a 2 is en...

Страница 26: ...tion the temperature entered is assumed to be positive After this first character entry there are three restrictions which apply prior to depres sion of the ENTER key and they are 1 A decimal point MAY be included as an entry It should be preceded by at least one number or a 0 2 The entered temperature must be within the operational temperature range of the bath 3 The entered actual temperature mu...

Страница 27: ...ay will be provided showing that timed reports are disabled and the present mode is polled 1 If the user wishes to operate in a point to point environment with or without timed interval reports a 2 should be entered now COMMUNICATION MODE 1 Polled 2 Point to Point Reports Disabled 1 If the TE 3000 was previously configured to operate in a point to point RS 232 environment the following display wil...

Страница 28: ...controller has a full duplex communication link RS 232 and can therefore issue these unrestricted timed reports If the bath is re configured in the polled mode of communication operations the controller has a half duplex communication link RS 485 and cannot issue timed reports The report option will be automatically disabled and a message indicating the report status will be displayed on the LCD s...

Страница 29: ...erface If this software is updated and a new release is issued by CANNON Instrument Company a diskette will be sent to the users along with instructions on how to perform this upgrade Download button A download push button is provided on the rear panel of the TE 3000 which places the instrument into the download mode When this mode is entered the front panel display will show the version number of...

Страница 30: ...scription of the Commands Queries Responses and Reports on this interface please refer to section D which follows D Commands queries responses and reports The protocol on the serial interfaces includes commands and queries which are sent by a master controlling computer and responses which are sent by the TE 3000 in reply to every command or query it receives The TE 3000 may also issue timed inter...

Страница 31: ...gh Tech Road State College PA 16803 USA Format examples Therefore the format for a typical COMMAND line is C And the format for a typical QUERY line is Q On the following pages is a table of Command types with possible arguments along with a description of the action taken by the TE 3000 and its subsequent response ...

Страница 32: ... o c s i h t f i e l p m a x e n a s A C D C 1 s i h t e k i l k o o l t h g i m e s n o p s e r e h T S E Y C D C 1 F D e n o n s i h T s t r o p e r l l a n i d n a D C L e h t n o t i e h n e r h a F s e e r g e d n i e r u t a r e p m e t l l a y a l p s i D A s e c n e r e f e r e r u t a r e p m e t l l a r o f e d o m t i e h n e r h a F e h t n i h t a b e h t s e c a l p d n a m m o c d n...

Страница 33: ... n o p s e r o n s i e r e h T B S C r o F e r u t a r e p m e t e e r g e d e l o h w d e i f i c e p s e h t r o f n i B t e s f f o n o i t a r b i l a c e h t s t e S s e e r g e d n i d e n i f e d s a t e s f f o e h t h t i w F r o C r o f d e n i f e d s a r o f s u n i m r o s u l p a h t i w g n o l a d e n i f e d e b t s u m n i b t i e h n e r h a F r o s u i s l e C e h T s u n i m r...

Страница 34: ... g n i e b s s e r d d a k r o w t e n d n a e c a p s e n o t s a e l t a y b d e w o l l o f s i e l a c s e r u t a r e p m e t e h t t a h w e t a c i d n i o t F r o f s t i g i d o w t n o l o c a y b d e w o l l o f s r u o h e t a c i d n i o t s t i g i d o w t r o e n o n e h t e h t e c n i s d e s p a l e e m i t e h t s i h c i h w s d n o c e s r o f s t i g i d o w t d n a l a m i c...

Страница 35: ...d be scheduled on a monthly basis Cleaning of painted surfaces and front panel The finish of the TE 3000 is a baked on epoxy enamel in colors of almond and black This paint is virtually immune to most solvents however if acetone is rubbed on the painted surfaces the paint will be removed A good quality household cleaner may be used on all painted surfaces and the front panel Do not spray the clean...

Страница 36: ... taken to not bend or crush the fins of the heat sinks Assemble the side panels back onto the Bath Vessel Housing by reversing the removal procedure Compressed air cleaning If a central or self contained air compressor line with a blow gun is not available compressed air aerosol cans may be obtained from most computer supply dealers These cans may be used in conjunction with a long nozzle to facil...

Страница 37: ...ng the inside of the bath vessel Depending on the grade of alcohol being used as a bath liquid the frequency of replacing the bath liquid may vary greatly The contami nants contained in solution with the alcohol will remain as sediment in the bath after the alcohol evaporates If the bath temperature is always maintained at a cold temperature and the holes at the top are always covered the rate of ...

Страница 38: ...34 CANNON TE 3000 Thermoelectric Constant Temperature Bath Version 1 1c May 2005 CANNON Instrument Company 2139 High Tech Road State College PA 16803 USA This page intentionally left blank ...

Страница 39: ...ot be liable for any special incidental or conse quential damages or any damage to plant personnel equipment or products directly or indirectly resulting from the use or misuse of any product Representations and warranties made by any person including dealers and representatives of the Company which are inconsistent in conflict with or in excess of the terms of this warranty shall not be binding u...

Страница 40: ...n __________ 2139 High Tech Road State College PA 16803 USA Please include the following Required information The Return Authorization number RA The name and telephone number of the person at your company to contact regarding the product Shipping and billing instructions for the return of the product to your location A detailed explanation of the reason for the return If the product is not covered...

Страница 41: ...owing parts represent those items in the TE 3000 which are believed to be either expendable or may exhibit a short lifetime Other parts possibly damaged by misuse or neglect may not be included in this list but may be obtained by contacting CANNON Instrument Company directly PART REPLACEMENT PART DESCRIPTION P25 2455 Fuse M 12A 250V 1 for 120V model P27 1300 Lamp Fluorescent 7W P28 0718 RTD Probe ...

Страница 42: ...38 CANNON TE 3000 Thermoelectric Constant Temperature Bath Version 1 1c May 2005 CANNON Instrument Company 2139 High Tech Road State College PA 16803 USA This page intentionally left blank ...

Страница 43: ...a e r g s i e r u t a r e p m e t k n i s t a e h r i A n e e b s a h t i n U s u i s l e C 5 3 n a h t e h t h t i w e m i t e m o s r o f g n i t a r e p o l l a w a t s n i a g a d e k c o l b s n a f g n i l o o c n o i t c u r t s b o r e h t o r o d n a r a e r e h t t a s n o i t c u r t s b o m o r f y a w a t i n u e v o M l o o c o t s k n i s t a e h w o l l A s e d i s e l b a n u s i ...

Страница 44: ... s s e V h t a B r e p p u e h t n o t e g r a t r o f d e t c e n n o c e b t o n d e e n e s o h r e h t o e h T s u i s l e C s e e r g e d 0 1 e v o b a s e r u t a r e p m e t t o n s i e b o r p e r u t a r e p m e T h t a b e h t n i d e l l a t s n i y l r e p o r p l e s s e v e b d l u o h s r o s n e s e r u t a r e p m e t e h t f o p i t m o t t o b e h T e l f f a b e t i h w e h t f...

Страница 45: ......