Platform/pump assembly
The platform and pump assembly may be completed with the following
items provided by
CANNON
®
Overflow Platform
Assembly screws (2)
Reciprocating Pump
Pre-cut tubing (approx. 12", 8" & 4" lengths)
1. If the rear platform is not yet attached to the Bath Vessel Housing,
attach it using the two 6-32 screws provided. The tabs which will
hold the Overflow Jar in place should be facing upward.
2. After the platform is securely in place, attach the 12" tubing to the
bath intake at the top rear of the Bath Vessel Housing by pushing the
tubing firmly over the intake nozzle. Then attach the other end of the
tube to the hose connection on top of the Reciprocating Pump (next
to the electric wire) in the same fashion.
3. Insert the other end of the reciprocating pump into the small hole on
the Overflow Platform so that the pump rests on the platform.
4. Attach one end of the
8" tube to the bottom
connection on the
Reciprocating Pump.
5. Attach the 4" tube to
the Bath Vessel
Housing overflow
outlet located above
the Overflow Plat-
form. Place the
Overflow Jar on the
platform, inserting
the other end of the
4" tube through the
lid of the jar.
6. Attach the other end of the 8" tube to the connector on the underside
of the Overflow Jar. Make sure that all connections are secure.
7. Ensure that the cable from the liquid pump is secured to the
PUMP
connection on the lower Control Chassis (see instructions, page 7).
Figure 4: Overflow jar with platform and
pump installed