background image

31

CANNON

®

 CT-1000 Constant Temperature Bath

Revision 3.1a—February, 1998; 

CANNON

®

 Instrument Company

P.O . Box 16 • State College, PA • USA

NBS MONOGRAPH 150: Repairing separated mercury
columns

The following text outlines procedures for repairing separated mercury
columns in thermometers. The text is adapted from:  

NBS MONOGRAPH 150 Liquid-In-Glass Thermometry
Wise, Jacquelyn A.

NOTE

Many inquiries are received concerning separated mercury column which
occur especially during shipment. Since no means of avoiding such
occurrences has yet been found, some directions for joining mercury may
be helpful and are described below.

(A)

The bulb of the thermometer may be cooled in a solution of common salt,
ice, and water (or other cooling agent) to bring the mercury down slowly
into the bulb. If the salt solution does not provide sufficient cooling,
carbon dioxide snow (dry ice) may be used. Since the temperature of dry
ice is approximately -78°C (-108°F), and mercury freezes at approxi-
mately -40°C (-40°F), the mercury will solidify. Cool only the bulb and
never the stem or mercury column. Moderate tapping of the bulb on a
rubber stopper or similar soft spongy object, or the application of centrifu-
gal force, by swinging the thermometer in a short arc (i.e. use of centrifu-
gal force), usually serves to unite the mercury in the bulb. Care must be
taken to warm the top of the bulb first, so pressures in the bulb due to
expanding mercury may be relieved.

(B)

If there is a contraction chamber above the bulb or an expansion chamber
at the top of the  thermometer, the mercury can sometimes be united by
warming the bulb until the column reaches the separated portions in either
enlargement. Great care is necessary to avoid filling the expansion
chamber completely with mercury, which might produce pressures large
enough to burst the bulb. (The expansion chamber should never be more
than 2/3 full). Joining the mercury is more readily accomplished if the
quantity in either cavity has been shattered into droplets by tapping the
thermometer laterally against the hand.

This procedure should not be used it if requires the thermometer to be
heated above 260°C (500°F) and the bulb should never be heated in an
open flame.

(C)

As a last resort, especially for thermometers having no expansion cham-
bers, small separated portions of the column can sometimes be dispersed
if mercury is warmed into droplets tiny enough to leave space for the gas
to by-pass. The thermometer is heated, and the droplets are collected by
the rising mercury column.

Organic liquid procedures

The procedure for thermometers containing organic liquids is similar.
Separated liquid in the stem can be vaporized and permitted to drain down

Содержание CT-1000

Страница 1: ...ting viscometer tubes thermometers 9 Filling the bath 10 Draining the Bath 11 3 BATH OPERATION 13 Turning on the CT 1000 13 Setting the bath temperature 14 Fixed temperature adjustments 14 Variable te...

Страница 2: ...PA USA F APPENDIX F THERMOMETRY 27 KINEMATIC VISCOSITY AND TEMPERATURE 27 ASTM THERMOMETER TABLES 28 ASTM D 445 CHECKING THE ICE POINT 29 NBS MONOGRAPH 150 REPAIRING SEPARATED MERCURY COLUMNS 31 G WAR...

Страница 3: ...bility and ease of use it is also suitable for any other application where temperatures must be maintained within close tolerances Temperature range The CANNON CT 1000 will maintain temperatures of 20...

Страница 4: ...ps provide glare free illumination of the bath A solid state control circuit equipped with a stainless steel encased thermistor provides proportional temperature control Amotor driven stirrer ensures...

Страница 5: ...ns who use the CT 1000 should follow these basic safety procedures Do not place the CT 1000 system on an unstable cart or stand The CT 1000 should be placed on a stable laboratory table or bench If an...

Страница 6: ...4 CANNON CT 1000 Constant Temperature Bath Revision 3 1a February 1998 CANNON Instrument Company P O Box 16 State College PA USA This page intentionally left blank...

Страница 7: ...st screwdriver Allen wrench seven hole covers rubber thermometer holder and instruction manual Box 2 contains the bath housing including the electronics drawer Box 2 also contains a smaller box with t...

Страница 8: ...ove the front top cover and rear top cover see Figure 4 and 5 Use caution when removing the rear top cover because the tempera ture control probes and heating elements are attached to it Glass jar ins...

Страница 9: ...binet 5 Lower the glass jar into the cabinet so it seats evenly on the rubber support ring see Figure 8 6 Unwrap the two glass panels Place the thinner of the two piece of glass in the slot closest to...

Страница 10: ...with the bath until the top of the jar forms a tight seal with the covers see Figure 10 Make sure you tighten the set screws uniformly so the jar remains level 15 Run the AC cord through the rear pane...

Страница 11: ...before filling the bath with fluid Inserting viscometer tubes thermometers The top cover of the CT 1000 contains seven apertures 51 mm 2 in diameter for the insertion of viscometer tube holders Two a...

Страница 12: ...to add fluid until the bath liquid level has risen to approxi mately 40 mm 1 5 of the top of the jar 4 Turn the instrument power ON and incrementally heat the bath to desired control temperature while...

Страница 13: ...le container to hold all of the liquid drained from the bath approximately 20 22 liters or 4 5 5 gallons Make sure that the bath liquid is within 10 C of ambient temperature Then insert a tube into th...

Страница 14: ...12 CANNON CT 1000 Constant Temperature Bath Revision 3 1a February 1998 CANNON Instrument Company P O Box 16 State College PA USA This page intentionally left blank...

Страница 15: ...ters to match their typeface on the panel Turning on the CT 1000 1 Turn the Power BATH switch on Figure 14 A NOTE The bath heaters will not be activated if there is no liquid in the bath Figure 14 Bat...

Страница 16: ...will glow continuously until the bath temperature is approximately that selected with the Temperature SELECT switch when the light will begin to blink The light above the Temperature Adjust control w...

Страница 17: ...nd allow several minutes for the bath to equilibrate NOTE If a desired bath temperature is within two or three degrees of one on the Temperature SELECT dial it is sometimes possible to attain the tem...

Страница 18: ...16 CANNON CT 1000 Constant Temperature Bath Revision 3 1a February 1998 CANNON Instrument Company P O Box 16 State College PA USA This page intentionally left blank...

Страница 19: ...Check for working motor by connecting to outlet with the proper voltage see motor label for correct voltage Check connections for sensors on rear panel Check temperature setting it must be above exis...

Страница 20: ...18 CANNON CT 1000 Constant Temperature Bath Revision 3 1a February 1998 CANNON Instrument Company P O Box 16 State College PA USA This page intentionally left blank...

Страница 21: ...r shaft run out and impeller blade alignment by placing the motor on a soft surface with impeller facing up Connect the motor to the appropriate AC power source and observe the impeller and shaft If r...

Страница 22: ...ame side of a shaft as the first mark and is about the same length the bending was not great enough to alter the condition and should be repeated with more force applied If the mark has shifted to the...

Страница 23: ...rovides a list of operating temperature ranges and some liquids suitable for use in these ranges C 0 0 2 o t C 0 0 1 S N O I T P O D I U L F H T A B e r u t a r e p m e T C e g n a R s d i u q i L h t...

Страница 24: ...ed Because in some cases there is a tendency for algae formation some degree of water treatment may be necessary Water can be used at temperatures close to the boiling point but water replenishment to...

Страница 25: ...other than one that is available on the Temperature SELECT switch is to be accessed frequently it may be possible to obtain this temperature by adjusting the trimpot that coincides with the Tempera t...

Страница 26: ...24 CANNON CT 1000 Constant Temperature Bath Revision 3 1a February 1998 CANNON Instrument Company P O Box 16 State College PA USA This page intentionally left blank...

Страница 27: ...10 LAMINATEDSAFETYGLASS P27 1320 PLATEGLASS P27 2190 LIGHTEDSWITCHROCKER120V P27 2220 PANELMOUNTLED P27 2230 TEMPERATURERESETSWITCH P27 2230 11 LAMPRESETSWITCH P27 2250 SOLIDSTATERELAY25A P27 2280 BAL...

Страница 28: ...26 CANNON CT 1000 Constant Temperature Bath Revision 3 1a February 1998 CANNON Instrument Company P O Box 16 State College PA USA This page intentionally left blank...

Страница 29: ...ter CANNON Instru ment Company recommends the use of a calibrated ASTM kinematic viscosity thermometer ASTM Thermometers Each ASTM kinematic viscosity thermometer measures only 3 degrees on a scale su...

Страница 30: ...vailable from CANNON Instrument Company ASTM CENTIGRADE THERMOMETERS TYPE CATALOGUE RANGE 74C 9311 K47 55 4 to 52 6 C 73C 9311 K45 41 4 to 38 6 C 126C 9311 K77 27 4 to 24 6 C 127C 9311 K81 21 4 to 18...

Страница 31: ...83 5 to 88 5 F 28F 9311 L10 97 5 to 102 5 F 46F 9311 L45 119 5 to 124 5 F 29F 9311 L17 127 5 to 132 5 F 47F 9311 L52 137 5 to 142 5 F 48F 9311 L59 177 5 to 182 5 F 129F 9311 L99 197 5 to 202 5 F 30F 9...

Страница 32: ...ewar vessel with the crushed ice and add enough distilled and preferably precooled water to form a slush Do not float the ice Place the thermometer into the slush packing the ice gently around the ste...

Страница 33: ...e of centrifu gal force usually serves to unite the mercury in the bulb Care must be taken to warm the top of the bulb first so pressures in the bulb due to expanding mercury may be relieved B If ther...

Страница 34: ...of the bulb by rotating the thermometer and tapping it against the palm of the hand When the entire surface has been washed rotate the bubble to the top of the bulb and reunite the mercury as describe...

Страница 35: ...damages or any damage to plant personnel equipment or products directly or indirectly resulting from the use or misuse of any product sold by the Company except as set forth in and limited by the fore...

Страница 36: ...2139 High Tech Road State College PA 16803 USA Please include the following Required information The Return Authorization number RA The name and telephone number of the person at your company to conta...

Отзывы: