17
CANNON
®
CCS-2100 with VISCPRO
®
Instruction & Operation Manual
Version 2.0a — November 2010;
CANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
Use the ID field to input instrument identification information using up to
30 alphanumeric characters.
The
S/N:
field should correspond to the four-digit serial number from the
label on the CCS rear service panel.
The probe constant values (a0, a1, and a2) must be identical to the values
on the calibration certificate provided with your temperature probe.
Verify that the values correspond.
Note that firmware and communication port settings are automatically
determined. These values cannot be altered from this window.
When you have verified all settings for all editable fields, click
OK
.
Viewing/editing other setup information
If your instrument has already been set up by a technician, you can use
the instructions in this section of the manual to check or, if necessary,
change the instrument settings.
1. Click
Configure
from the VISCPRO
®
menu bar.
2. Select your instrument from the list of available instruments.
3. Select
Tray (Test) Settings
from the list of configuration options.
The
Tray (Test) Settings
window will appear.
The Tray (Test) Settings window