![C&H M 2 2 0 Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/candh/m-2-2-0/m-2-2-0_user-manual_3070834025.webp)
A-1
APPENDIX A: CONNECTORS
MAXIMUM VOLTAGE/CURRENT. The maximum voltage
that may be applied to any connector on the M220 is 200
VDC, 125 VAC rms, or 175 VAC peak--these limits apply
only if the product is installed in a humidity-controlled
(<60% RH) environment where airborne contaminants and
transients are controlled, and there is NOT a relay
connection made to power mains. If these conditions
CANNOT be maintained, then the maximum voltage is 60
VDC, 48 VAC-rms or 68 VAC-peak.
The maximum current (non-inductive) that may be applied to
the M220 is:
Per Switch: 2 ADC, 2 AAC peak
Per Module: 8 ADC, 8 AAC peak
Pin #
Use
Pin #
Use
Pin#
Use
1
CH0_HI
16
CH3_HI
31
CH3_LO
2
CH0_LO
17
CH2_HI
32
CH2_LO
3
CH7_LO
18
CH1_HI
33
CH1_LO
4
CH6_LO
19
CH7_HI
34
GND
5
CH5_LO
20
CH6_HI
35
GND
6
CH4_LO
21
CH5_HI
36
GND
7
MUXA_LO_COM
22
CH4_HI
37
GND
8
MUXB_HI_COM
23
MUXA_HI_COM
38
GND
9
CH8_LO
24
MUXB_LO_COM
39
GND
10
CH9_LO
25
CH8_HI
40
GND
11
CH10_LO
26
CH9_HI
41
GND
12
CH11_LO
27
CH10_HI
42
CH15_LO
13
CH12_HI
28
CH11_HI
43
CH15_HI
14
CH13_LO
29
CH12_LO
44
CH14_LO
15
CH13_HI
30
CH14_HI
Figure A-1. Front Panel I/O Signals
C&H Technologies, Inc. <> 445 Round Rock West Drive <> Round Rock, Texas 78681 <> www.chtech.com
Содержание M 2 2 0
Страница 10: ...4 C H Technologies Inc 445 Round Rock West Drive Round Rock Texas 78681 www chtech com...
Страница 24: ...18 C H Technologies Inc 445 Round Rock West Drive Round Rock Texas 78681 www chtech com...
Страница 26: ...A 2 C H Technologies Inc 445 Round Rock West Drive Round Rock Texas 78681 www chtech com...
Страница 27: ...N O T E S C H Technologies Inc 445 Round Rock West Drive Round Rock Texas 78681 www chtech com...
Страница 28: ...C H Technologies Inc 445 Round Rock West Drive Round Rock Texas 78681 www chtech com...