Installation Procedure
Candela Corporation
Page 6 of 8
Candela Corporation Proprietary
8503-01-0821, Revision A
Handpiece Assembly Installation
Refer to figure 1 for handpiece assembly system connection locations.
Figure 2 Delivery System and Handpiece Assembly
A- Handpiece Control Connector
B- Cryogen Output Connector
C- Fiber Proximal End
D- Glass Windows
E- Distance Gauge
F- O Ring
1.
Remove the handpiece assembly from the accessory kit package.
2.
With the system turned off, place the distal end of the handpiece assembly into
the calport.
3.
Connect the handpiece control (round, electrical plug) connector to the handpiece
control receptacle on the front of the laser system.
4.
Connect the cryogen output connector to the cryogen output receptacle on the
front of the laser system as follows: Using both hands, push the knurled
connector on the front of the laser system away from you while gently pushing the
Insert distance
gauge here
A
B
C
D
F
E