34
Code 3: Advance limit switch fault.
If the error occurs repeatedly during normal use contact Technical Support for
advice.
Code 4: Table limit switch fault.
Both limit switches flagged. Contact Technical Support for advice.
Code 5: Advance drive overload.
If the blade has collided with an obstruction, restart the instrument and carefully
drive the blade away from the obstruction.
If the error occurs repeatedly during normal use contact Technical Support for
advice.
Code 6: Oscillating head fault.
Restart the instrument. If the problem persists contact Technical Support for
advice.
Code 7: Oscillating head fault.
If motion of the vibrating head has been obstructed, restart the instrument and
carefully drive the blade away from the obstruction.
If the error occurs repeatedly during normal use contact Technical Support for
advice.
Code 8: Instrument overheat fault.
Contact Technical Support for advice.
Code 9: Memory Fault.
Restart the instrument. If the problem persists contact Technical Support for
advice.
Code 10: Alignment tool communications error.
If the alignment tool has been unplugged during the blade alignment
procedure, restart the instrument.
Code 12: Upper Table limit switch fault.
Upper limit switch flags continually or intermittently. Contact Technical Support
for advice.
Code 13: Lower Table limit switch fault.
Upper limit switch flags continually or intermittently. Contact Technical Support
for advice.
If the error occurs repeatedly during normal use contact Technical Support for
advice.
Содержание 7000smz
Страница 26: ...26 Correct any problems and press SLICE to retry...