You  must  now  choose  which  type  of 
blade  you  want  to  calibrate.  Scroll  to 

the  appropriate  blade  using  the 

 

or 

 keys and press 

‘MENU’ to select. 

Press 

‘SLICE’  to  continue.  The 

instrument  will  now  re-position  the 
blade  with  respect  to  the  alignment 
device ready for adjustment. 

 

 
 

The  display  will  now  change  to  that 
shown  on  the  left.  Note  that  the 
uppermost  line  shows  the  current  Z 
axis (vertical) deflection. At this point it 
indicates  zero  because  the  blade  is 
not moving. 

 
Press 

‘SLICE’ as instructed and when the blade is vibrating note the Z axis deflection 

reading. 
Press 

‘SLICE’ again to stop the blade vibration.  

 
Loosen  the  lower  locking  screw  just  sufficiently  tight  to  retain  its  grip  on  the  blade 
holder but without locking it in place. 
The blade holder can be tilted in a clockwise direction by turning the adjusting screw 
clockwise using the supplied hexagonal driver. Tighten the lower screw. 
Conversely, the blade holder can be tilted in a counter-clockwise direction by turning 
the adjusting screw counter-clockwise. 
 
Press 

‘SLICE’ to set the blade vibrating again and note the Z axis deflection. Stop the 

blade vibrating. If the error has increased you should repeat the adjustment but in the 
opposite direction. 
 
Continue  the  above  process  until  the  Z  axis  deflection  has  been  reduced  to 
acceptable limits. Once the blade is within acceptable limits the lower locking screw 
may be tightened to a torque of 30-40 cN.m. Do not over-tighten the screw 
Recheck the blade alignment. Repeat the adjustment procedure if required. 
 
 
When you have completed the alignment procedure press the 

‘RETURN’ key and 

follow the removal instructions as shown on the display. 
 
In certain circumstances, for example if a steel blade is fitted and you have specified 
a ceramic blade, the control system may position the blade incorrectly and not detect 
the blade edge. In these instances the display will show the following message: 

 

Содержание 7000smz

Страница 1: ...1 www campdeninstruments com INSTRUCTION MANUAL Vibrating Blade Tissue Slicer NOVEMBER 2019 DCN 2389 Document Ref 7000 v2 4 Eng...

Страница 2: ...xis of oscillation 6 10 Settings 7 Options 7 1 Light source 7 2 Magnifier 7 3 Microscope mounting point 8 Maintenance and Service 9 Returns Decontamination Form 10 Technical support Error codes 11 Pac...

Страница 3: ...uid The 7000smz instrument benefits in this area by being built on a heavy rigid cast base giving exceptional immunity to secondary vibration transmission The purpose of this manual is to allow the us...

Страница 4: ...at maximum speed and only after the uncut tissue has been lowered away from the blade eliminating the possibility of the blade being dragged back across the uncut tissue causing damage Dynamic feedbac...

Страница 5: ...An adjustable tissue platen is available as an optional extra The tissue bath is also located and held in place within the outer bath by magnets Inner and outer tissue bath background with standard ti...

Страница 6: ...6 A lower power 2x magnifying lens assembly can be fitted as an alternative to the microscope 7000smz fitted with magnifying glass and cold light source...

Страница 7: ...to an acute cutting edge Because of the relative softness of stainless steel for optimum performance it is recommended that stainless steel blades should be used once only or at a maximum changed ever...

Страница 8: ...rinsed with clean water on a regular basis The specimen platen and the inner stainless steel bath as shown in the illustration below are autoclavable The specimen holder is located by a circular magn...

Страница 9: ...be set for your particular voltage supply The voltage is set by prising out the fuse holder drawer and re inserting it such that the voltage legend for your supply is aligned with the mark on the inl...

Страница 10: ...is the normal operating display Press the MENU key again and the display will change to that shown Use the or keys to scroll to the line SETTINGS Then scroll to the line EDIT USER NAME and select Use...

Страница 11: ...changes to the oscillating frequency amplitude and section thickness will be remembered under that user name so that when that user comes back to the instrument and selects that user name those settin...

Страница 12: ...eys can now be used to change the parameter s value Press the MENU key to confirm and store the change Frequency settings can be changed in 5Hz steps and amplitude can be changed in 0 25mm steps Note...

Страница 13: ...ove the bath move the clamp lever to the left and pull the bath away from the instrument Pressing LOAD BATH keypad will cause the table to rise to a preset nominal position The display will then rever...

Страница 14: ...Section 6 2 and scroll to the EXIT MENU line and press the MENU key If necessary load a bath to the table and press the LOAD BATH key The table will raise the bath to a nominal position and the opera...

Страница 15: ...ey to reset the height datum Press the HEIGHT key once more to exit Bringing the specimen to the start position Cutting a slice Press the ADVANCE key and using the rotary knob to control the advance s...

Страница 16: ...d then raised once more by the thickness of the section in readiness for another cut When the cutting head has been retracted to the start position the section thickness will be highlighted on the dis...

Страница 17: ...via the manual slice procedure the display will indicate NO SLICE PROFILE HAS BEEN CREATED This display will time out and revert to the basic manual operating screen When a satisfactory cut window has...

Страница 18: ...ection 6 10 If you try to operate the instrument in the automatic mode without having previously set a cutting profile via the manual slice procedure the display will indicate NO SLICE PROFILE HAS BEE...

Страница 19: ...ow the instructions on screen to CONTINUE or ABORT the procedure Pressing MENU at this stage will allow access to menu functions permitting changes to frequency and amplitude etc Note that the EMERGEN...

Страница 20: ...ment has been halted by use of the Emergency Stop facility it must be reset before operations can continue Twist the red Emergency Stop button clockwise to unlatch it When the button is unlatched the...

Страница 21: ...designed so that the blade may be safely handled without the effective cutting length of the blade edge being exposed to possible damage or injuring the operator After carefully unwrapping the blade...

Страница 22: ...trument this incorporates a magnet and can simply be fitted over the blade with the magnet holding it in position See the following illustration This will provide some protection to the user and reduc...

Страница 23: ...ecured by two screws to the oscillating faceplate A third screw locks the blade holder in position once the alignment procedure has been completed See the illustration below Fit the blade holder as sh...

Страница 24: ...in its optimal position relative to the calibration device for best alignment evaluation To enter the alignment routine press the LOAD BATH key and hold for 2 seconds The bath table will be moved to...

Страница 25: ...hexagonal driver Tighten the lower screw Conversely the blade holder can be tilted in a counter clockwise direction by turning the adjusting screw counter clockwise Press SLICE to set the blade vibra...

Страница 26: ...26 Correct any problems and press SLICE to retry...

Страница 27: ...r keys Press the MENU key to confirm the change Use the or keys to scroll to line EXIT MENU and press the MENU key to get back to the basic menu display 2 When in an automatic slice mode the park posi...

Страница 28: ...strated below The light source can be switched on and the light intensity controlled via a combined press button on the body of the light source where the flexible arms are mounted A brief press will...

Страница 29: ...should then be fixed to the top of the instrument cover using the screws provided Do not over tighten the screws This is illustrated below If you also have the optional light source the magnifier sho...

Страница 30: ...cope free from vibration The Campden CL200 cold light source see above can be fitted directly to the top of the 7000smz 2 independently of the microscope For ease of access you should fit the light so...

Страница 31: ...t be accepted for service or repair unless the unit has been adequately and properly packaged Additionally instruments will not be accepted without prior authorisation and have been certified as being...

Страница 32: ...eria fungi etc YES NO 2 Radioactive Sources YES NO IF YES DO NOT RETURN 3 Chemical Hazard mercury salts acids bases etc YES NO 4 Rabies BSE CJD nCJD etc YES NO 5 Any Other Hazard PLEASE SPECIFY YES NO...

Страница 33: ...rial number Build date Blade run time Software revision code Any error codes that you may have witnessed This information may be readily accessed from the basic menu screen as follows Use the or keys...

Страница 34: ...n obstructed restart the instrument and carefully drive the blade away from the obstruction If the error occurs repeatedly during normal use contact Technical Support for advice Code 8 Instrument over...

Страница 35: ...35...

Страница 36: ...Specimen holder 1 off Opti Cal optical Calibration unit 1 off Stainless steel blades 10 off 7550 1 SS Ceramic blades 2 off 7550 1 C Cross head Philips PoziDriv screwdriver 1 off Hexagonal drivers 2 o...

Страница 37: ...nt adjustable specimen holder 7000 4 2 Blade holders 5000mz Blade holder For 7000smz prior to serial number 7000smz 020 and 5000mz prior to s n 5000mz 1004 7000 5 1 Blade holder special angle 7000 5 2...

Страница 38: ...cutting head 37 mm nominal Blade oscillation frequency 7000smz Minimum 50 Hz Maximum 120 Hz amplitude dependent Frequency step size 5 Hz Blade oscillation amplitude 7000smz Minimum 0 5 mm nominal Max...

Страница 39: ...ibility Directive 89 336 EEC The Low Voltage Directive 73 23 EEC 2 UK Regulations The Supply of Machinery Safety Regulations 1992 SI 1992 3073 Electricity at Work Regulations 1989 3 European Standards...

Отзывы: