25
You must now choose which type of
blade you want to calibrate. Scroll to
the appropriate blade using the
or
keys and press
‘MENU’ to select.
Press
‘SLICE’ to continue. The
instrument will now re-position the
blade with respect to the alignment
device ready for adjustment.
The display will now change to that
shown on the left. Note that the
uppermost line shows the current Z
axis (vertical) deflection. At this point it
indicates zero because the blade is
not moving.
Press
‘SLICE’ as instructed and when the blade is vibrating note the Z axis deflection
reading.
Press
‘SLICE’ again to stop the blade vibration.
Loosen the lower locking screw just sufficiently tight to retain its grip on the blade
holder but without locking it in place.
The blade holder can be tilted in a clockwise direction by turning the adjusting screw
clockwise using the supplied hexagonal driver. Tighten the lower screw.
Conversely, the blade holder can be tilted in a counter-clockwise direction by turning
the adjusting screw counter-clockwise.
Press
‘SLICE’ to set the blade vibrating again and note the Z axis deflection. Stop the
blade vibrating. If the error has increased you should repeat the adjustment but in the
opposite direction.
Continue the above process until the Z axis deflection has been reduced to
acceptable limits. Once the blade is within acceptable limits the lower locking screw
may be tightened to a torque of 30-40 cN.m. Do not over-tighten the screw
Recheck the blade alignment. Repeat the adjustment procedure if required.
When you have completed the alignment procedure press the
‘RETURN’ key and
follow the removal instructions as shown on the display.
In certain circumstances, for example if a steel blade is fitted and you have specified
a ceramic blade, the control system may position the blade incorrectly and not detect
the blade edge. In these instances the display will show the following message: