6. Specifications
Sensor element:
Measurement Specialties Micro-BetaCHIP 10K3MCD1
thermistor
Survival range
Thermistor:
–50 to 100 °C
Overmolded joint and cable:
–50 to 70 °C
Measurement range:
–40 to 70 °C
Time constant:
31 s in still air
7.5 s in a wind speed of 3 m/s
0.5 s in rolling water or antifreeze
Maximum cable length:
305 m (1000 ft)
Accuracy
1
Worst case:
±0.60 °C (–40 to 70 °C)
±0.49 °C (–20 to 70 °C)
Maximum submergence:
45 m (150 ft) (63 psi)
Interchangeability error:
±0.60 °C at –40 °C
±0.38 °C at 0 °C
±0.10 °C at 25 °C
±0.30 °C at 50 °C
±0.45 °C at 70 °C
Steinhart-Hart equation error:
≤ 0.02 °C (–40 to 70 °C)
Stainless-steel sheath
Diameter:
0.16 cm (0.063 in)
Length:
5.84 cm (2.3 in)
Overmolded joint
Diameter:
1.02 cm (0.40 in)
Length:
4.24 cm (1.67 in)
Cable:
Santoprene®, 0.220 in diameter
Cable/probe connection:
ATUMTM heat shrink
109SS Temperature Probe
5