TDR100
;Apply Topp's 3rd order polynomial:
8: Polynomial (P55)
1: 1
Reps
2: 4
X Loc [ WC_CH3 ]
3: 4
F(X) Loc [ WC_CH3 ]
4: -0.053 C0
5: .292
C1
6: -0.055 C2
7: 0.0043 C3
8: 0.0
C4
9: 0.0
C5
;Measure Electrical Conductivity on the same CS610 TDR probe connected to Channel 3.
9: TDR100 Measurement (P119)
1: 0
SDM
Address
2: 3
Electrical
Conductivity
3: 3001
MMMP Mux & Probe Selection
4: 4
Waveform
Averaging
5: 1
Vp
6: 250
Points
7: 9.5
Cable
Length
(meters)
8: 5
Window
Length
(meters)
9: .3
Probe
Length
(meters)
10: .085
Probe
Offset
(meters)
11: 5
Loc [ EC_CH3 ]
12: 1.74
Mult
;units = Siemens/meter
13: 0.0
Offset
;Get the raw waveform for the CS610 probe on Channel 3.
;When creating this program in Edlog, the user must reserve 259 input locations
;for data storage. How? In Edlog click on
;"Edit", select "Input Labels". In the Input Location Editor click on "Edit", select
;"Insert Block", and identify 259 input storage location for the captured waveform.
;Note: 9 parameter el 250 waveform data points = 259 input location/waveform.
10: TDR100 Measurement (P119)
1: 0
SDM
Address
2: 1
Waveform
(256
locs)
3: 3001
MMMP Mux & Probe Selection
4: 4
Waveform
Averaging
5: 1
Vp
6: 250
Points
7: 9.5
Cable
Length
(meters)
8: 5
Window
Length
(meters)
9: .3
Probe
Length
(meters)
10: .085
Probe
Offset
(meters)
11: 11
Loc [ WF_1 ]
12: 1.0
Mult
13: 0.0
Offset