SECTION 2.  STORING DATA FROM CR10, 21X, AND CR7

2-3

2.1.2  USE OF TWO STORAGE MODULES WITH

21X OR CR7

For large volumes of data, two SMs can be
connected to the 21X or CR7.  Use SMCOM to
set one SM as Fill and Stop, and the other as
Ring Memory.

2.2  *8 AND *9 MODE -- KEYBOARD

INITIATED DATA TRANSFER

2.2.1  *8 MODE - CR10

The *8 Mode is used to manually initiate a Final
Storage Data transfer from the CR10 to an
external peripheral.  Data transfer rate is 9600
baud.  Refer to the discussion of the *8 Mode
(Section 4.2) in the CR10 Operator's Manual for
full details.

2.2.2  *9 MODE - 21X OR CR7

The F*9 Mode in the 21X and CR7 dataloggers
is used to transfer a specified block of data from
Final Storage to the Storage Module.  Consult
the appropriate datalogger manual for details.
This function is different from the *9 Mode in the
CR10.

2.3  STORING BURST MEASUREMENT

DATA FROM THE CR10 OR 21X

The BURST Measurement is a special
datalogger instruction used to make high rate
analog voltage measurements.  Two options
are available for data retrieval with this
instruction:  1) Data are stored in datalogger
Input Storage Locations or 2) Raw data are sent
out the datalogger serial port.  In the second
case, the Storage Module can capture the data
at 76,800 baud.

BURST data is output by the datalogger in a
unique format which includes calibration
measurements at the beginning and end of the
series of measurements.  SPLIT will translate
this BURST data.  (Section 4.3.1 in the PC208
manual.)

2.3.1  BURST DATA WITH CR10

The CR10 requires a special PROM to use the
BURST Measurement Instruction (23).  Keying
a "3" in the second digit from the right in
Parameter 04 of Instruction 23 selects the
Storage Module as the destination of the
BURST data.  All SM's connected should be set
to Fill and Stop and have consecutive
addresses.  The BURST data will be sent to the
first available (lowest addressed) Storage
Module, followed by the next lowest addressed,
and so on.

2.3.2  BURST DATA WITH 21X

To send BURST Measurement (Instruction 23)
data to the Storage Module from the 21X, set
the Destination option (Parameter 04) to 2 for
Serial port output at 76,800 baud.

Содержание SM192

Страница 1: ...SM192 716 STORAGE MODULE INSTRUCTION MANUAL REVISION 6 14 93 COPYRIGHT c 1989 1993 CAMPBELL SCIENTIFIC INC...

Страница 2: ...rchantability or fitness for a particular purpose CAMPBELL SCIENTIFIC INC is not liable for special indirect incidental or consequential damages Products may not be returned without prior authorizatio...

Страница 3: ...FROM CR10 21X AND CR7 2 1 2 1 Instruction 96 Output Data Under Program Control 2 1 2 1 1 Multiple Storage Modules with CR10 2 2 2 1 2 Use of Two Storage Modules with 21X or CR7 2 3 2 2 8 and 9 Mode K...

Страница 4: ...nd Retrieve Datalogger Programs 6 1 LIST OF TABLES 6 1 1 D Mode Keyboard Commands for the CR10 and 21X 6 1 A 1 9 Commands for Storage Module A 1 LIST OF FIGURES 1 2 1 Storage Module and SC12 Cable 1 2...

Страница 5: ...ers released after February 1989 and CR7 dataloggers released after August 1991 contain the OSX and OS7 Series PROMS which support two major SM functions that earlier versions do not 1 The D storage a...

Страница 6: ...the datalogger or from data retrieval interfaces such as the SC532 9 pin Peripheral to RS232 Interface the SM232A Storage Module RS232 Interface or the PC201 Clock SIO Tape Read Card Pin 1 of the 9 p...

Страница 7: ...connected and sends the backlogged data the next time it executes Instruction 96 The SM can store data either under program control Instruction 96 or under user control 8 Mode with CR10 9 Mode with 2...

Страница 8: ...SECTION 1 SM192 716 STORAGE MODULE OVERVIEW 1 4 Figure 1 4 1 Operation with Current CR10 21X and CR7 Figure 1 4 2 Printer Enable Operations Figure 1 4 3 Telecommunications Command State...

Страница 9: ...in the data from SMCOM or the CR10 9 Mode Command 3 The Storage Module will not store back to back file marks if a File Mark is the last thing stored in memory a second file mark will not be stored 1...

Страница 10: ...a to Storage Module When output to the Storage Module is enabled with Instruction 96 the Storage Module s may either be left on site and periodically exchanged or brought to the site to milk the data...

Страница 11: ...this manual respectively The default Storage Module address on reset as sent from the factory is 1 This will be sufficient for most users At a given time only one Storage Module will execute commands...

Страница 12: ...og voltage measurements Two options are available for data retrieval with this instruction 1 Data are stored in datalogger Input Storage Locations or 2 Raw data are sent out the datalogger serial port...

Страница 13: ...t When in this setting the SM can automatically change from capturing data at 76 800 baud to capturing data at 9600 baud This feature allows the use of two different baud rates in a single program The...

Страница 14: ...to the Storage Module automatically when it is properly configured and connected to the CR21 Use SCOM to set the Storage Module baud rate to 300 to collect CR21 data The Encode Decode switch should be...

Страница 15: ...3H command is used only during testing The Dump Pointer is an internal Storage Module pointer used for keeping track of the current start of dump for Storage Module to Storage Module data dumps and fo...

Страница 16: ...6 800 baud Bytes from the Storage Module input buffer are permanently stored in byte pairs The Campbell Scientific Inc dataloggers Final Storage Format data are always grouped in pairs Printable ASCII...

Страница 17: ...Campbell Scientific Inc s SC12 9 pin ribbon cable links the Storage Module to the SC532 Figure 5 1 1 The SC532 requires 12 VDC nominal which is usually furnished by plugging the AC DC adaptor unit in...

Страница 18: ...SECTION 5 DATA RETRIEVAL 5 2 5 1 1 Storage Module Connected With SC532 5 1 2 Storage Module Connected to PC201 Card 5 1 3 Communication Via the CR10 5 1 4 Storage Module Connected With SM232A...

Страница 19: ...1 c Execute SMCOM PC208 manual 5 3 2 NON IBM COMPATIBLES The following instructions apply to a configuration using a terminal emulator program and the SM232A It is assumed for this example that the t...

Страница 20: ...n data retrieval program 5 3 3 CR10 TELECOMMUNICATIONS INTERFACE Make the connections Figure 5 1 3 and establish communications with the CR10 Send the command XM CR to the CR10 where X is the Storage...

Страница 21: ...Display Description D 13 00 Enter D Mode 7XA 7X 00 Address Storage Module X 1YA Store Program in SM as Y Y 1 8 2YA Load Program Y from SM 3YA Erase Program Y from SM 13 0000 command completed The pro...

Страница 22: ...1 minute with the SM192 or 4 minutes with the SM716 the number of good RAM chips will be displayed 01 06 6 RAM chips in SM192 01 22 22 RAM chips in SM716 Most commands have at least one response Adva...

Страница 23: ...y location pointer to SRP 9 Oldest data to SRP 1 5 will loop within file boundaries 0 7 9 allows display to cross boundaries If no file is found the display returns 07 00 At this point another file ca...

Страница 24: ...A 3 COMMAND DISPLAY DESCRIPTION 9 DISPLAY ADDRESSES OF CONNECTED SM S XXXXXXXX 1 occupied 0 unoccupied 87654321 Addresses 8 1 from left to right 10 CHANGE ADDRESS 10 0X X is current address enter yA...

Страница 25: ...B 1 APPENDIX B SAMPLE FLOW DIAGRAM FOR DATA RETRIEVAL PROGRAM...

Страница 26: ...EVICES The Printer Enable Method Section 4 is used for non datalogger SM applications The PE line must be low when power is applied or removed If a switch is used to enable data storage the switch mus...

Страница 27: ...DC and Pin 7 CLK HS is low 0 VDC the processor is set to receive data asynchronously on Pin 9 TD When the PE line is dropped data remaining in the input buffer are stored and the location pointers are...

Страница 28: ...stored number of good Memory chips number of Errors logged max 255 number of Available storage locations number of locations Full Storage Reference Pointer Display Location Pointer and Checksum sum of...

Страница 29: ...P is advanced to the beginning of the next file The DLP is advanced as the SM reads through the data until a File Mark or the Storage Reference Pointer is encountered For long files the delay involved...

Страница 30: ...ch When used with a CR10 or 21X this switch can be set to 1 8 The address has no effect when used with other Campbell Scientific Inc dataloggers Following reset the SM address defaults to 1 b is the B...

Отзывы: