Met One 034B Wind Set
11
The expression for wind speed (U) is:
U = MX + B
where
M = multiplier
X = number of pulses per second (Hertz)
B = offset
lists the multipliers (M) and offsets (Off) to obtain met
re
s/second
or miles/hour when the
Pulse Count
instruction is configured to output the
result in Hz.
TABLE 7-2. Wind Speed Multiplier*
Model
Met
re
s/Second
Miles/Hour
034B
M = 0.7989
Off = 0.28
M = 1.787
Off = 0.63
*When configured to output counts, the multiplier above is divided by the
execution interval in seconds.
7.4.2 Wind Direction
The CR200(X) dataloggers use the
ExDelSE()
instruction to measure wind
direction. All other CRBasic dataloggers use the
BRHalf()
instruction (see
(p. B-1)
, for example).
Excitation voltages, range codes, and multipliers for our dataloggers are listed
. The multiplier value converts the sensor’s millivolt output to
degrees. Appendix
Wind Direction Measurement Theory
(p. D-1)
, has
additional information on the measurement instructions.
TABLE 7-3. Parameters for Wind Direction
CR300
CR800
CR1000
CR5000
CR3000
CR200(X)
CR6
Measurement
Range, Integration
2500 mV,
60 Hz,
reverse
excitation
5000 mV,
60 Hz,
reverse
excitation
N/A
5000 mV,
60 Hz,
reverse
excitation
Excitation Voltage
2500 mV
5000 mV
2500 mV
2500 mV
Multiplier
720
720
0.288
720
Offset
0
0
0
0
7.4.3 Wind Vector Processing Instruction
The
Wind Vector
output instruction is used to process and store mean wind
speed, unit vector mean wind direction, and standard deviation of the wind
direction (optional) from the measured wind speed and direction values.