background image

LWS Dielectric Leaf Wetness Sensor 

7.2  Wiring 

TABLE 7-1.  Wire Color, Wire Function, and Data Logger Connection 

Wire 

Color 

Wire Function 

Data Logger Connection Terminal  

Brown or 

White 

Voltage 

excitation input 

configured for voltage excitation

1

EX, 

VX

 (voltage excitation)

 

Orange or 

Red 

Analog voltage 

output 

configured for single-ended analog 

input

1

SE

 (single-ended, analog input)  

Clear 

Shield 

 (analog ground) 

1

terminals are automatically configured by the measurement instruction.  

 

7.3  Programming 

Short Cut

 is the best source for up-to-date data logger programming code.  

If your data acquisition requirements are simple, you can probably create and 

maintain a data logger program exclusively with 

Short Cut

. If your data 

acquisition needs are more complex, the files that 

Short Cut 

creates are a great 

source for programming code to start a new program or add to an existing 

custom program. 

Short Cut 

cannot edit programs after they are imported and edited 

in 

CRBasic Editor

Short Cut 

tutorial is available in Section 

4

QuickStart

 

(p. 1)

If you wish to 

import 

Short Cut 

code into 

CRBasic Editor 

to create or add to a customized 

program, follow the procedure in Appendix Appendix 

A

Importing Short Cut 

Code Into CRBasic Editor

 

(p. A-1)

Programming basics for CRBasic data 

loggers are provided in the following sections. Complete program examples for 

CRBasic data loggers can be found in Appendix 

B

Example Programs

 

(p. B-1)

Programming basics and programming examples for Edlog data loggers are 

provided a

www.campbellsci.com/old-manuals

. 

7.3.1  Voltage Measurement 

The LWS requires excitation voltage between 2.5 and 5 Vdc. It produces an 

output voltage dependent on the dielectric constant of the medium surrounding 

the sensor. Output voltage ranges from 10 to 50% of the excitation voltage.  

Except for the CR200(X), CRBasic data loggers use the 

BRHalf() 

instruction 

to measure the sensor output. The 

BRHalf() 

instruction and parameters are as 

follows: 

BRHalf

(Dest,Reps,Range,SeChan,ExChan,MeasPEx,ExmV,RevEx,Settling, 

fN1/Integ,Mult,Offset) 

The CR200(X) uses the 

ExDelSE()

 CRBasic instruction to measure the sensor 

output. 

The ExDelSE() instruction 

and parameters are as follows: 

ExDelSE

( Dest, Reps, SEChan, ExChan, ExmV, Delay, Mult, Offset 

NOTE 

Содержание LWS

Страница 1: ...LWS Dielectric Leaf Wetness Sensor Revision 11 18 Copyright 2008 2018 Campbell Scientific Inc...

Страница 2: ...ucts to CSI CSI will return such Products by surface carrier prepaid within the continental United States of America To all other locations CSI will return such Products best way CIP port of entry per...

Страница 3: ...ing address is CAMPBELL SCIENTIFIC INC RMA _____ 815 West 1800 North Logan Utah 84321 1784 For all returns the customer must fill out a Statement of Product Cleanliness and Decontamination form and co...

Страница 4: ...beginning work Wear a hardhat and eye protection and take other appropriate safety precautions while working on or around tripods and towers Do not climb tripods or towers at any time and prohibit cl...

Страница 5: ...s 4 7 Installation 5 7 1 Field Installation 5 7 2 Wiring 6 7 3 Programming 6 7 3 1 Voltage Measurement 6 7 3 2 Minutes Dry Minutes Wet or Contaminated and Minutes Wet 7 7 3 3 Interpreting Data 7 8 Mai...

Страница 6: ...le of Contents ii Tables B 1 CR1000X Example Program Wiring B 1 B 2 CR6 Example Program Wiring B 2 CRBasic Examples B 1 CR1000X Program for Measuring the LWS B 1 B 2 CR6 Program for Measuring the LWS...

Страница 7: ...fic Although the LWS is rugged it should be handled as a precision scientific instrument Over time the accumulation of dust and bird droppings can cause the dry output to rise We recommend that the se...

Страница 8: ...le Sensors and Devices box type LWS You can also locate the sensor in Sensors Miscellaneous Sensors folder Double click LWS Dielectric Leaf Wetness Sensor Enter the Dry threshold mV and Wet threshold...

Страница 9: ...the output of the sensor in LoggerNet PC400 RTDAQ or PC200W to make sure it is making reasonable measurements Short Cut uses the execution interval to make the minutes wet dry and contaminated calcul...

Страница 10: ...spectrum but selectively reject much of the energy in the near infrared The surface coating of the LWS absorbs well in the near infrared region but the white color reflects most of the visible radiati...

Страница 11: ...2 Wiring p 6 and Section 7 3 Programming p 6 Short Cut does this work for you See Section 4 QuickStart p 1 for a Short Cut tutorial 7 1 Field Installation The LWS includes two holes for mounting the...

Страница 12: ...in Section 4 QuickStart p 1 If you wish to import Short Cut code into CRBasic Editor to create or add to a customized program follow the procedure in Appendix Appendix A Importing Short Cut Code Into...

Страница 13: ...minutes wet can then be totaled and stored for any given period table interval Minutes wet or contaminated can be considered a wet condition or a contaminated condition depending on the user s evaluat...

Страница 14: ...debris will cause the dry output to increase and change the Boolean threshold Clean the sensing surface with a moist cloth periodically or when elevated dry output is detected The LWS leaf wetness se...

Страница 15: ...Cut program following the procedure in Section 4 QuickStart p 1 Finish the program On the Advanced tab click the CRBasic Editor button The program opens in CRBasic with the name noname CR_ Provide a n...

Страница 16: ...00X Brown or White Excitation VX1 Orange or Red Analog Out SE1 Clear Analog Ground CRBasic Example B 1 CR1000X Program for Measuring the LWS CR1000X Declare Variables and Units Public BattV Public PTe...

Страница 17: ...f LWmV 274 Then LWMDry 0 08333333 Else If LWmV 284 Then LWMWet 0 08333333 Else LWMCon 0 08333333 EndIf EndIf Call Data Tables and Store Data CallTable Hourly CallTable Daily NextScan EndProg B 2 Examp...

Страница 18: ...urement BattV Battery BattV Default Wiring Panel Temperature measurement PTemp_C PanelTemp PTemp_C 60 LWS Dielectric Leaf Wetness Sensor measurement LWmV BrHalf LWmV 1 mV5000 U2 U1 1 2500 False 10000...

Страница 19: ...Asia Location Bangkok Thailand Email info campbellsci asia Website www campbellsci asia China Location Beijing P R China Email info campbellsci com cn Website www campbellsci com cn Spain Location Ba...

Отзывы: