Table 12-2: COMM Status LED Activity
Green
Flashing
Module has been configured by data logger and is receiving sync
signals
Orange
Flashing
Device has not been configured by data logger
Red
Single-Flash Device has been configured but is not receiving sync signals
Double-
Flash
Device has encountered a scan timeout
Solid
Error
12.3 CPI port LEDs
Green and Orange LEDs on the CPI bus indicate the current operation of the CPI bus as shown in
the following table.
Table 12-3: CPI port LED activity
Green
Flashing
Frame sync
Orange
Flash
CPI activity
Solid
CPI bus fault
12.4 Replacing a GRANITE measurement
module
An existing VOLT 100 series or CDM-A100 series can be replaced with a new module using the
same CPI address without requiring a change to the program.
VOLT 100 series
40