Appendix C. EC155 Sample Cell and Intake Maintenance
C-2
4.
Pull the bypass tube out of the bypass tube connection sleeve.
5.
Disconnect the heater cable connection. See FIGURE
6.
Loosen the four screws that secure the intake mounting plate in place and
remove the plate. See FIGURE
7.
Locate the black nut next to the stainless steel block of the sample cell that
holds the stainless steel sample tubing in place by compressing two rubber
o-rings around the tubing. Loosen this nut roughly two full turns but don’t
remove it. See FIGURE
8.
Carefully pull the intake away from the sample cell, sliding the stainless
steel sample tubing from the sample cell. See FIGURE
FIGURE C-1. Identification of parts for detaching intake with heater
cable shown disconnected
FIGURE C-2. Removal of the vortex intake
Mounting Plate Screw
(1 of 4)
Compression Nut
Heater Cable Connector
Intake Mounting Plate
Содержание EC155
Страница 78: ...E 1 Appendix E Safety Data Sheets SDS E 1 Molecular Sieve Type 13X ...
Страница 79: ...Appendix E Safety Data Sheets SDS E 2 ...
Страница 80: ...Appendix E Safety Data Sheets SDS E 3 ...
Страница 81: ...Appendix E Safety Data Sheets SDS E 4 ...
Страница 82: ...Appendix E Safety Data Sheets SDS E 5 E 2 Magnesium Perchlorate ...
Страница 83: ...Appendix E Safety Data Sheets SDS E 6 ...
Страница 84: ...Appendix E Safety Data Sheets SDS E 7 ...
Страница 85: ...Appendix E Safety Data Sheets SDS E 8 ...
Страница 86: ...Appendix E Safety Data Sheets SDS E 9 ...
Страница 87: ...Appendix E Safety Data Sheets SDS E 10 ...
Страница 88: ...Appendix E Safety Data Sheets SDS E 11 ...
Страница 89: ...Appendix E Safety Data Sheets SDS E 12 E 3 Decarbite ...
Страница 90: ...Appendix E Safety Data Sheets SDS E 13 ...
Страница 91: ...Appendix E Safety Data Sheets SDS E 14 ...
Страница 92: ...Appendix E Safety Data Sheets SDS E 15 ...
Страница 93: ...F 1 Appendix F EC155 Packing Information ...