background image

CS615 WATER CONTENT REFLECTOMETER

7

The calibration coefficients in the example
programs are demonstrative only.  See Section
9 for information on calibration.

10.1  SAMPLE PROGRAM 1

Simple program using the Period Averaging
Instruction (P27) of a CR10 datalogger to read a
single CS615. The measurement is taken
hourly and the period and water content are
written to output storage.

CR10

CS615

Single-Ended

Channel 1 (SE1)

green

Control Port 5 (C5)

orange

The red lead is connected to 12VDC and the
black and shield are connected to ground.

;{CR10}
;
*Table 1 Program

01:

5.0

Execution Interval (seconds)

1:  If time is (P92) ;take reading hourly

1:

0000

Minutes (Seconds --) into a

2:

60

Interval (same units as above)

3:

30

Then Do

2:  Do (P86)

1:

45

Set Port 5 High;enable CS615

3:  Period Average (SE) (P27)

1:

1

Reps

2:

4

2 V Peak to Peak/200 kHz
Max. Freq.

3:

1

SE Channel

4:

10

No. of Cycles

5:

5

Timeout (units = 0.01 seconds)

6:

1

Loc [ 615period ]

7:

.001

Mult

8:

0.0

Offset

4:  Do (P86) ;disable CS615

1:

55

Set Port 5 Low

5:  Polynomial (P55) ;convert period to water content

1:

1

Reps

2:

1

X Loc [ 615period ]

3:

2

F(X) Loc [ 615water  ]

4:

-0.187

C0

5:

0.037

C1

6:

0.335

C2

7:

0.0

C3

8:

0.0

C4

9:

0.0

C5

6:  Do (P86)

1:

10

Set Output Flag High

7:  Real Time (P77)

1:

0220

Day,Hour/Minute

8:  Sample (P70)

1:

2

Reps

2:

1

Loc [615period]

9:  End (P95)

End Program

10.2  SAMPLE PROGRAM 2

Simple program using the Pulse Count Instruction
(P3) of a CR10, CR500 or 21X datalogger to read
a single CS615.  The Water Content
Reflectometer is queried when Flag 1 is set high.
The measurement is made and the period and
water content values are written to output storage.

CR10 or 21X

CS615

Pulse Count

Channel 1 (P1)

green

Control Port 5 (C5)

orange

The red lead is connected to 12VDC and the
black and shield are connected to ground.

;{CR10, CR500 or 21X}
;
*Table 1 Program

01:

1.0

Execution Interval (seconds)

1:  If Flag/Port (P91) ;set flag1 high to initiate
reading

1:

11

Do if Flag 1 is High

2:

30

Then Do

2:  Do (P86) ;enable CS615

1:

45

Set Port 5 High

3:  Beginning of Loop (P87) ;delay for complete
count interval

1:

1

Delay

2:

2

Loop Count

4:  End (P95)

5:  Pulse (P3) ;determine CS615 output frequency

1:

1

Reps

2:

1

Pulse Input Channel

3:

21

Low Level AC, Output Hz

4:

1

Loc [ 615kHz    ]

5:

.001

Mult ;convert to kHz

6:

0.0

Offset

Содержание CS615

Страница 1: ...CS615 WATER CONTENT REFLECTOMETER INSTRUCTION MANUAL version 8221 07 REVISION 10 96 COPYRIGHT c 1995 1996 CAMPBELL SCIENTIFIC INC ...

Страница 2: ... merchantability or fitness for a particular purpose CAMPBELL SCIENTIFIC INC is not liable for special indirect incidental or consequential damages Products may not be returned without prior authorization To obtain a Returned Materials Authorization RMA contact CAMPBELL SCIENTIFIC INC phone 435 753 2342 After an applications engineer determines the nature of the problem an RMA number will be issue...

Страница 3: ...IONS 4 1 Accuracy 2 4 2 Resolution 2 4 3 Operating Range 2 5 INSTALLATION 5 1 Orientation 3 5 2 Potential Problems with Improper Insertion 3 6 WIRING 4 7 DATALOGGER INSTRUCTIONS 7 1 Introduction 4 7 2 Pulse Count 4 7 3 Period Measurement 5 8 MAINTENANCE 5 9 CALIBRATION 9 1 General 5 10 SAMPLE PROGRAMS 6 10 1 Sample Program 1 7 10 2 Sample Program 2 7 10 3 Sample Program 3 8 10 4 Sample Program 4 9...

Страница 4: ... of two stainless steel rods connected to a printed circuit board A shielded four conductor cable is connected to the circuit board to supply power enable the probe and monitor the pulse output The circuit board is encapsulated in epoxy High speed electronic components on the circuit board are configured as a bistable multivibrator The output of the multivibrator is connected to the probe rods whi...

Страница 5: ...content but it is also affected by electrical conductivity Free ions in soil solution provide electrical conduction paths which result in attenuation of the signal applied to the waveguides This attenuation both reduces the amplitude of the high frequency signal on the probe rods and affects the shape of the oscillating signal The attenuation reduces oscillation frequency at a given water content ...

Страница 6: ...s were performed at various water contents and over the temperature range from 10 C to 30 C The calibration information presented in Section 9 is for a temperature of 20 C The following equation can be used to interpolate the temperature coefficient for a range of volumetric water content θv values Coeftemperature v v 3 46 10 0 019 0 045 4 2 θ θ To apply this correction the following equation can ...

Страница 7: ... in the calibration for water content The Pulse Count instruction of a CR10 CR500 21X or CR7 dataloggers can be used with the CS615 output connected to a pulse count channel The Period Measurement instruction of the CR10 or CR500 can be used with the CS615 output connected to a single ended analog channel 7 2 PULSE COUNT It is important to understand the event sequence during the Instruction 3 Pul...

Страница 8: ...uction to enable the CS615 and these instructions can take a significant amount of time to execute During the second time through the Loop Instruction the pulse counters see the CS615 output for the full execution interval At the end of this interval the accumulators transfer to RAM and are reset When the program execution then moves to the Pulse Count Instruction P3 the value in RAM is converted ...

Страница 9: ...fect on the calibration but the magnitude is dependent on the clay type electrical conductivity dS m 1 calibration 1 0 θ τ τ τ v 0187 0 037 0 335 2 1 8 θ τ τ τ v 0 207 0 097 0 288 2 3 0 θ τ τ τ v 0 298 0 361 0 096 2 θv is the volumetric water content on a fraction basis i e 0 20 is 20 volumetric water content τ is the CS615 output period in milliseconds 9 2 CALIBRATION FOR A SPECIFIC SOIL The cali...

Страница 10: ...nvert period to water content 1 1 Reps 2 1 X Loc 615period 3 2 F X Loc 615water 4 0 187 C0 5 0 037 C1 6 0 335 C2 7 0 0 C3 8 0 0 C4 9 0 0 C5 6 Do P86 1 10 Set Output Flag High 7 Real Time P77 1 0220 Day Hour Minute 8 Sample P70 1 2 Reps 2 1 Loc 615period 9 End P95 End Program 10 2 SAMPLE PROGRAM 2 Simple program using the Pulse Count Instruction P3 of a CR10 CR500 or 21X datalogger to read a single...

Страница 11: ...ion to volumetric water content is invoked using the Polynomial Instruction The water content values are written to output storage Attention to program structure when using the Pulse Count Instruction with sensors that are periodically enabled is necessary to ensure accurate results See Section 7 2 for a detailed description of the Pulse Count Instruction CAUTION 1 The probe rods of the CS615 are ...

Страница 12: ...w 17 Do P86 1 10 Set Output Flag High 18 Sample P70 1 48 Reps 2 7 Loc water 1 19 End P95 End Program 10 4 SAMPLE PROGRAM 4 Program using the Period Averaging Instruction P27 of CR10 datalogger and AM416 multiplexer to read 48 CS615 probes CAUTION 1 The probe rods of the CS615 are essentially antennae which transmit and receive radio waves Interference can occur when enabled probes are in close pro...

Страница 13: ...hannel 4 10 No of Cycles 5 5 Timeout units 0 01 seconds 6 1 Loc msec 1 7 001 Mult 8 0 0 Offset 8 End P95 set ports enabling mux and probes low 9 Do P86 1 51 Set Port 1 Low 10 Do P86 1 53 Set Port 3 Low apply calibration 11 Polynomial P55 1 48 Reps 2 1 X Loc Period 1 3 49 F X Loc WatCont 1 4 0 187 C0 5 0 037 C1 6 0 335 C2 7 0 0 C3 8 0 0 C4 9 0 0 C5 12 Do P86 1 10 Set Output Flag High 13 Real Time P...

Страница 14: ...C3 P1 P2 P3 21x Using Pulse Count Reset Clock H1 L1 H2 L2 H1 L1 H2 L2 H1 L1 H2 L2 H1 L1 H2 L2 1 2 16 enable signal enable signal enable signal enable signal enable signal enable signal enable signal 48 6 5 4 3 2 1 CR10 or 21X AM416 CS615 SE1 SE2 SE3 CR10 Using Period Measurement signal 46 signal 47 signal 48 SE Single Ended Input Channel ...

Отзывы: