CS451/CS456 Submersible Pressure Transducer
7.3.1 SDI-12
TABLE 7-6. SDI-12 Wiring
Color
CS451/CS456
Function
CR800
CR5000
CR3000
CR1000
CR200(X)
Series
CR510
CR23X
CR10X
Red +12VDC 12V
12V
Black Power
Ground
G
G
G
White
SDI-12 Signal
*Control Port
C1/SDI-12
Control Port
Blue
GND GND G G
Yellow
GND GND G G
Clear Shield
GND
G
G
*dedicated SDI-12 port on CR5000
7.3.2 RS-232
TABLE 7-7. RS-232 Wiring
Color
CS451/CS456
Function
Connection
RS-232 9-pin
Red +12vdc
Power
Source
Black
Power Ground
Power Ground
White
RS-232 Tx (Output)
Transmit
Pin 2 Rx (Input)
Blue
RS-232 Rx (Input)
Receive
Pin 3 Tx (Output)
Yellow
Digital Ground
Ground
Pin 5 GND
Clear Shield
GND
Ground
7.4 Programming
SCWin is the preferred method for programming the datalogger.
SCWin generates a wiring diagram that shows how to connect
the CS451/CS456 to your Campbell Scientific datalogger. This
section is for users who are using CRBasic or Edlog to program
their datalogger.
NOTE
The sections that immediately follow are for CRBasic and Edlog.
NOTE
Keyboard/Display users and SCWin users can jump ahead to the Maintenance
section.