![Campbell CS451 Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/campbell/cs451/cs451_product-manual_3793103024.webp)
CS451/CS456 Submersible Pressure Transducer
12
7.3.1 SDI-12 Data Logger Connections
TABLE 7-1. Wire Colo
u
r, Function, and Data Logger Connection
for SDI-12 Measurements
Wire Colo
u
r
Wire Function
Data Logger Connection
White
SDI-12 Signal
C
terminal
1
or
U
configured for SDI-12
2
Clear
Shield
⏚
(analog
ue
ground)
Red
Power
12V
,
Black
Power Ground
G
Blue
Ground
⏚
(analog
ue
ground)
Yellow
Ground
⏚
(analog
ue
ground)
1
Dedicated SDI-12 port on CR5000.
2
U
terminals are automatically configured by the measurement instruction.
For the CR6 and CR1000X data loggers, triggering conflicts may occur when a
companion terminal is used for a triggering instruction such as
TimerInput()
,
PulseCount()
, or
WaitDigTrig()
. For example, if the CS451 is connected to
C3 on a CR1000X, C4 cannot be used in the
TimerInput()
,
PulseCount()
, or
WaitDigTrig()
instructions.
7.3.2 RS-232 Connections
TABLE 7-2. Wire Colo
u
r, Function, Data Logger
Connection, and RS-232 Pin
Wire
Colo
u
r
CS451/CS456
Function
Data Logger
Connection
RS-232 9-Pin
Red
+12vdc
12V
,
Black
Power Ground
G
White
RS-232 Tx (Output)
Odd-numbered
C
terminal
Pin 2 Rx (Input)
Blue
RS-232 Rx (Input)
Even-numbered
C
terminal
Pin 3 Tx (Output)
Yellow
Digital Ground
⏚
(analog
ue
ground)
Pin 5 GND
Clear
Shield GND
⏚
(analog
ue
ground)
7.4 Programming
Short Cut
is the best source for up-to-date data logger programming code.
If data acquisition requirements are simple, a data logger program can probably
be created and maintained exclusively with
Short Cut
. If your data acquisition
needs are more complex, the files that
Short Cut
creates are a great source for