CS450/CS455 Pressure Transducer
5.2 RS-232
TABLE 5-2. RS-232 Wiring
Color
CS450/CS455
Function
Connection
RS-232 9-pin
Red +12vdc
Power
Source
Black
Power Ground
Power Ground
White
RS-232 Tx (Output)
Transmit
Pin 2 Rx (Input)
Blue
RS-232 Rx (Input)
Receive
Pin 3 Tx (Output)
Yellow
Digital Ground
Ground
Pin 5 GND
Clear Shield
GND
Ground
5.3 A150/CWS900
The A150 desiccated case allows the CS450/CS455 to be connected to the
CWS900 wireless sensor and then used in a wireless network. FIGURE 5-1
shows the CS450 wired to the A150’s terminal block. The 26972 cable is
included with the A150.
BLUE
RED
CLEAR
GREEN
1
2
BLACK
3
4
WHITE
5
6
BLACK
#26972 Cable
#19520
CWS900 Connector
WHITE
BLUE
BLACK
YELLOW
CLEAR
RED
TUBE
N.C.
N.C.
Terminates in A150 Enclosure
A150
Pressure Sensor
CS450
Sensor TX (Out)
Sensor RX (In)
Power GND
Digital GND
Shield
Power
Vent
(Pin #)
FIGURE 5-1. A150 wired for CS450 pressure sensor and CWS900
connector
6. Programming
6.1 Using SCWIN
SCWIN is the easiest and typically the preferred method for programming the
datalogger. SCWIN generates a wiring diagram that shows how to connect the
CS450/CS455 to your Campbell Scientific datalogger.
The sections that immediately follow are for CRBasic and Edlog.
NOTE
16