![Campbell CS140 Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/campbell/cs140/cs140_product-manual_3811056027.webp)
Table 6-3: GET command transmitted data
Example
Description
0x02
1
STX
GET
GET command
:
Delimiting character
0
Address based on sensor ID
:
Delimiting character
0
Reserved for future use; zero default
:
Delimiting character
2C67
Checksum
:
Delimiting character
0x03
1
ETX
0x0D
1
Carriage return
1
These values are shown in hexadecimal format not ASCII.
Example of a
GET
request:
GET : 0 : 0 : 2C67 :
Example data returned by the
GET
command:
GET : 0 : 0 : 2C67 :
0 0 2 1000 0 60 0 2 1 1 0 0 0 1 7.0 0 0 10000 626C
Table 6-4: GET command returned data
Example
Description
0x02
1
STX
0
Sensor ID
0
Serial port protocol
2
Serial baud rate
1000
Sensor serial number (read only)
0
Luminance units
60
Continuous mode output interval
CS140 Background Luminance Sensor
21