Instruction Manual
23
Figure 9-2. Internal switches
Internal switch functions
Switch number
Function
4
Reserved for future use, set to OFF.
3
When switched to the ON position and the sensor
is power cycled this switch temporarily sets the
sensor communications port to a default RS-232
communication state at 38400 baud. This is useful
during field tests or maintenance when the sensor
has been remotely configured for RS-485 mode or
a baud rate your PC does not support. This change
is temporary and will not be stored to flash.
However, if the menu system is accessed and a
‘Save and exit’ command is performed these new
data rate settings will be committed to flash.
Once this switch is returned to its OFF position
and the sensor is power cycled the sensor will
return to its previous communications settings.
2
Reserved for future use, set to OFF.
1
When switched to the ON position this switch
will reset the sensor to its factory default values.
This reset will affect all communication settings.
This will take immediate effect upon power up.
NOTE: To use this the power supply must be
stable. Do not leave this switch set permanently.
Содержание CS120A
Страница 2: ...ii ...
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 12: ...vi ...
Страница 73: ...A 1 Appendix A CS120A CS125 block diagram ...
Страница 74: ...CS120A and CS125 Visibility and Present Weather Sensors A 2 ...
Страница 76: ...CS120A and CS125 Visibility and Present Weather Sensors B 2 ...
Страница 90: ......
Страница 93: ...Appendix D Present Weather Codes D 3 ...