User Guide
33
The calibration is performed from menu item 3 on the main terminal screen.
Please refer to the menu section (Section 13) of this manual for further
information on how to access this menu.
Once you have selected menu item 3 you should be presented with the following
screen.
Select option 1 to start the calibration. You will then be asked to confirm that you
would like to perform a calibration. Please note, once you have entered yes at this
point you will not be able to exit until the test is complete. However, power
cycling the unit at this point will have no adverse effect on the sensor.
Once you have started the tests you will be asked for the CS120A calibrator serial
number and coefficient with a confirmation at each step giving you the chance to
correct typing mistakes.
When asked for confirmation you do not need to press return after
you type ‘y’.
NOTE
CS120A CALIBRATION - MENU 3
ID 0
S/N 1009
(1) Perform calibration
(2) Restore the factory calibration
(9) Refresh
(0) Return to main menu
->
Starting calibration.
Input the CS120A calibrator serial number ->E2002
Is E2002 correct? (Y/N)?
Input the CS120A calibrator constant ->28.8
Is 28.8 correct? (Y/N)?
Place one calibration bung into each hood, then
press any key.
Do you want to perform a calibration Y/N?